WebGraph-Thermo-Hygrometer-MIB DEFINITIONS ::= BEGIN ---------------------------------------------------------------- -- Revison: 1.26 ---------------------------------------------------------------- -- 03.08.2006 wtWebGraphThermHygroSensorNo korrigiert -- ---------------------------------------------------------------- -- Revison: 1.25 ---------------------------------------------------------------- -- 28.06.2006 New: Community String Trap -- ---------------------------------------------------------------- -- Revison: 1.24 ---------------------------------------------------------------- -- 26.07.2005 #57606, WebGraph Thermo-Hygrometer -- CRON Zeitfunktion ---------------------------------------------------------------- -- Revison: 1.23 ---------------------------------------------------------------- -- 14.07.2005 #57606, WebGraph Thermo-Hygrometer -- FTP Funktionen ---------------------------------------------------------------- -- Revison: 1.22 ---------------------------------------------------------------- -- 17.05.2005 #57606, WebGraph Thermo-Hygrometer -- Trap NMS annotations ---------------------------------------------------------------- -- Revison: 1.21 ---------------------------------------------------------------- -- 19.04.2005 #57606, WebGraph Thermo-Hygrometer -- Mail Authentication ---------------------------------------------------------------- -- Revison: 1.20 ---------------------------------------------------------------- -- 05.11.2004 #57606, WebGraph Themo-Hygrometer -- Neu: Syslog -- ---------------------------------------------------------------- -- Revison: 1.19 ---------------------------------------------------------------- -- 08.09.2004 #57606 WebGraph Thermo-Hygrometer -- Neu: -- wtWebGraphThermHygroAlarmRHMin -- wtWebGraphThermHygroAlarmRHMax -- wtWebGraphThermHygroAlarmRHHysteresis -- wtWebGraphThermHygroPortSensorSelect -- wtWebGraphThermHygroAlarmDeltaTemp -- wtWebGraphThermHygroAlarmAHMin -- wtWebGraphThermHygroAlarmAHMax ---------------------------------------------------------------- -- Revison: 1.18 ---------------------------------------------------------------- -- 21.07.2004 #57601/3/4, Webio Digital -- Neu: -- Diagnosis ---------------------------------------------------------------- -- Revison: 1.17 ---------------------------------------------------------------- -- 28.06.2004 #57630/1/2, Webio Digital -- Neu: Puls Mode ---------------------------------------------------------------- -- Revison: 1.16 ---------------------------------------------------------------- -- 16.04.2004 neu: #57632, Webio Digital EA24x24 -- -- #57630/1, Webio Digital EA12x12 -- Neu: -- wtWebioEA12x12InputCounterClear ---------------------------------------------------------------- -- Revison: 1.15 ---------------------------------------------------------------- -- 27.02.2004 #57630/1, Webio Digital EA12x12 -- Neu: -- L�nge erweitert auf 32 Bit: -- wtWebioEA12x12InputValue -- wtWebioEA12x12OutputValue -- wtWebioEA12x12LoadControlView -- wtWebioEA12x12LCShutDownView -- wtWebioEA12x12SetOutput -- wtWebioEA12x12AlarmInputTrigger -- wtWebioEA12x12AlarmOutputTrigger -- wtWebioEA12x12PortOutputSafetyState ---------------------------------------------------------------- -- Revison: 1.14 ---------------------------------------------------------------- -- 16.02.2004 #57630/1, Webio Digital EA12x12 -- Neu: -- Texte: Location, Contact -- SNMP: ColdStart, WarmStart, ComunityString Read/RW -- Alarm: ColdStart / WarmStart trigger -- -- #57601, 1xWebThermometer -- #57603, 2xWebThermometer -- #57604, 8xWebThermometer -- Neu: -- Alarm Hysterese -- Alarm Clear Traps und Mails ---------------------------------------------------------------- -- Revison: 1.13 ---------------------------------------------------------------- -- 05.02.2004 #57601, 1xWebThermometer -- #57603, 2xWebThermometer -- #57604, 8xWebThermometer -- Neu: -- Texte: Location, Contact -- SNMP: ColdStart, WarmStart, ComunityString Read/RW -- Alarm: ColdStart / WarmStart trigger ---------------------------------------------------------------- -- Revison: 1.12 ---------------------------------------------------------------- -- 05.11.2003 #57601, 1xWebThermometer -- #57603, 2xWebThermometer -- #57604, 8xWebThermometer -- Neu: -- Mail Option: CSV attachment -- Offsetkalibrierung -- Alarm: Tcp Client mode ---------------------------------------------------------------- -- Revison: 1.11 ---------------------------------------------------------------- -- 28.10.2003 #57630/1, Webio Digital EA12x12 -- Neue Feature: -- Alarm : TCP Client mode -- INPUTS: Pulsfilter ---------------------------------------------------------------- -- Revison: 1.10 ---------------------------------------------------------------- -- 04.07.2003 #57630, Webio Digital EA12x12 -- Neuer Men�baum, neue Bin�rmodi -- ---------------------------------------------------------------- -- Revison: 1.09 ---------------------------------------------------------------- -- 24.03.2003 Trap Enterprise (Type 1, 2) korrigiert -- ---------------------------------------------------------------- -- Revison: 1.08 ---------------------------------------------------------------- -- 04.03.2003 Abgleich der Sprachversionen DE/US -- ---------------------------------------------------------------- -- Revison: 1.07 ---------------------------------------------------------------- -- 25.02.2003 Korrektur US: wtWebioEA12x12LoadControl ---------------------------------------------------------------- -- Revison: 1.06 ---------------------------------------------------------------- -- 24.02.2003 Neu: -- #57630, Webio EA12x12, LoadControl Group -- Korrektur: Trap variable Namen ---------------------------------------------------------------- -- Revison: 1.05 ---------------------------------------------------------------- -- 21.01.2003 Neu: -- #57630, Webio EA12x12 ---------------------------------------------------------------- -- Revison: 1.04 ---------------------------------------------------------------- -- 02.01.2003 Neu: -- #57601, 1xWebThermometer -- #57603, 2xWebThermometer, 8 Alarme ---------------------------------------------------------------- -- Revison: 1.03 ---------------------------------------------------------------- -- 21.11.2002 Neu: -- #57604, 2xWebThermometer, BOOTP, StartupSite -- #57603, 8xWebThermometer, BOOTP, StartupSite ---------------------------------------------------------------- -- Revison: 1.02 ---------------------------------------------------------------- -- 26.07.2002 Neu: -- #57604, 2xWebThermometer -- #57603, 8xWebThermometer -- IMPORTS OBJECT-TYPE FROM RFC-1212 TRAP-TYPE FROM RFC-1215 DisplayString, PhysAddress FROM RFC1213-MIB enterprises, TimeTicks, IpAddress, NetworkAddress FROM RFC1155-SMI; -- IpAddress ::= -- [APPLICATION 0] -- IMPLICIT OCTET STRING (SIZE (4)) -- Counter ::= -- [APPLICATION 1] -- IMPLICIT INTEGER (0..4294967295) -- Gauge ::= -- [APPLICATION 2] -- IMPLICIT INTEGER (0..4294967295) -- TimeTicks ::= -- [APPLICATION 3] -- IMPLICIT INTEGER (0..4294967295) -- DisplayString ::= OCTET STRING -- PhysAddress ::= OCTET STRING -- internet OBJECT IDENTIFIER ::= { iso org(3) dod(6) 1 } -- privat OBJECT IDENTIFIER ::= { internet 4 } -- enterprises OBJECT IDENTIFIER ::= { privat 1 } -- Wiesemann & Theis Com-Server MIB wut OBJECT IDENTIFIER ::= { enterprises 5040 } wtComServer OBJECT IDENTIFIER ::= { wut 1 } wtWebio OBJECT IDENTIFIER ::= { wtComServer 2 } -- wtWebGraph Tree wtWebGraphThermHygro OBJECT IDENTIFIER ::= { wtWebio 9 } -- wtWebGraphThermHygro Tree wtWebGraphThermHygroTemp OBJECT IDENTIFIER ::= { wtWebGraphThermHygro 1 } wtWebGraphThermHygroSessCntrl OBJECT IDENTIFIER ::= { wtWebGraphThermHygro 2 } wtWebGraphThermHygroConfig OBJECT IDENTIFIER ::= { wtWebGraphThermHygro 3 } wtWebGraphThermHygroDiag OBJECT IDENTIFIER ::= { wtWebGraphThermHygro 4 } -- wtWebGraphThermHygroConfig Tree wtWebGraphThermHygroDevice OBJECT IDENTIFIER ::= { wtWebGraphThermHygroConfig 1 } wtWebGraphThermHygroPorts OBJECT IDENTIFIER ::= { wtWebGraphThermHygroConfig 2 } wtWebGraphThermHygroManufact OBJECT IDENTIFIER ::= { wtWebGraphThermHygroConfig 3 } -- wtWebGraphThermHygroDevice Tree wtWebGraphThermHygroText OBJECT IDENTIFIER ::= { wtWebGraphThermHygroDevice 1 } wtWebGraphThermHygroTimeDate OBJECT IDENTIFIER ::= { wtWebGraphThermHygroDevice 2 } wtWebGraphThermHygroBasic OBJECT IDENTIFIER ::= { wtWebGraphThermHygroDevice 3 } wtWebGraphThermHygroDatalogger OBJECT IDENTIFIER ::= { wtWebGraphThermHygroDevice 4 } wtWebGraphThermHygroAlarm OBJECT IDENTIFIER ::= { wtWebGraphThermHygroDevice 5 } -- wtWebGraphThermHygroTimeDate Tree wtWebGraphThermHygroTimeZone OBJECT IDENTIFIER ::= { wtWebGraphThermHygroTimeDate 1 } wtWebGraphThermHygroTimeServer OBJECT IDENTIFIER ::= { wtWebGraphThermHygroTimeDate 2 } wtWebGraphThermHygroDeviceClock OBJECT IDENTIFIER ::= { wtWebGraphThermHygroTimeDate 3 } -- wtWebGraphThermHygroBasic Tree wtWebGraphThermHygroNetwork OBJECT IDENTIFIER ::= { wtWebGraphThermHygroBasic 1 } wtWebGraphThermHygroHTTP OBJECT IDENTIFIER ::= { wtWebGraphThermHygroBasic 2 } wtWebGraphThermHygroMail OBJECT IDENTIFIER ::= { wtWebGraphThermHygroBasic 3 } wtWebGraphThermHygroSNMP OBJECT IDENTIFIER ::= { wtWebGraphThermHygroBasic 4 } wtWebGraphThermHygroUDP OBJECT IDENTIFIER ::= { wtWebGraphThermHygroBasic 5 } wtWebGraphThermHygroSyslog OBJECT IDENTIFIER ::= { wtWebGraphThermHygroBasic 6 } wtWebGraphThermHygroFTP OBJECT IDENTIFIER ::= { wtWebGraphThermHygroBasic 7 } -- wtWebGraphThermHygroTemp Group ------------------------ wtWebGraphThermHygroSensors OBJECT-TYPE SYNTAX INTEGER (1..2) ACCESS read-only STATUS mandatory DESCRIPTION "Anzahl der Sensoren" ::= { wtWebGraphThermHygroTemp 1 } -- Interface Table wtWebGraphThermHygroSensorTable OBJECT-TYPE SYNTAX SEQUENCE OF WtWebGraphThermHygroSensorEntry ACCESS not-accessible STATUS mandatory DESCRIPTION "Liste der Sensoren." ::= { wtWebGraphThermHygroTemp 2 } wtWebGraphThermHygroSensorEntry OBJECT-TYPE SYNTAX WtWebGraphThermHygroSensorEntry ACCESS not-accessible STATUS mandatory DESCRIPTION "Nummern der Sensoren." INDEX { wtWebGraphThermHygroSensorNo } ::= { wtWebGraphThermHygroSensorTable 1 } WtWebGraphThermHygroSensorEntry ::= SEQUENCE { wtWebGraphThermHygroSensorNo INTEGER } wtWebGraphThermHygroSensorNo OBJECT-TYPE SYNTAX INTEGER (1..2) ACCESS read-only STATUS mandatory DESCRIPTION "Nummer des Sensors." ::= { wtWebGraphThermHygroSensorEntry 1 } -- wtWebGraphThermHygroTempValue Table wtWebGraphThermHygroTempValueTable OBJECT-TYPE SYNTAX SEQUENCE OF WtWebGraphThermHygroTempValueEntry ACCESS not-accessible STATUS mandatory DESCRIPTION "Liste Sensormesswerte korrespondiernd zu den Kanalnummern." ::= { wtWebGraphThermHygroTemp 3 } wtWebGraphThermHygroTempValueEntry OBJECT-TYPE SYNTAX WtWebGraphThermHygroTempValueEntry ACCESS not-accessible STATUS mandatory DESCRIPTION "Messwerte der Sensoren." INDEX { wtWebGraphThermHygroSensorNo } ::= { wtWebGraphThermHygroTempValueTable 1 } WtWebGraphThermHygroTempValueEntry ::= SEQUENCE{ wtWebGraphThermHygroTempValue OCTET STRING } wtWebGraphThermHygroTempValue OBJECT-TYPE SYNTAX OCTET STRING (SIZE (5)) ACCESS read-only STATUS mandatory DESCRIPTION "Messwert des Sensors, ASCII, 5 Zeichen lang." ::= { wtWebGraphThermHygroTempValueEntry 1 } -- wtWebGraphThermHygroBinaryTempValue Table wtWebGraphThermHygroBinaryTempValueTable OBJECT-TYPE SYNTAX SEQUENCE OF WtWebGraphThermHygroBinaryTempValueEntry ACCESS not-accessible STATUS mandatory DESCRIPTION "Liste bin�rer Messwerte korrespondiernd zu den Kanalnummern." ::= { wtWebGraphThermHygroTemp 4 } wtWebGraphThermHygroBinaryTempValueEntry OBJECT-TYPE SYNTAX WtWebGraphThermHygroBinaryTempValueEntry ACCESS not-accessible STATUS mandatory DESCRIPTION "Bin�re Messwerte der Sensoren." INDEX { wtWebGraphThermHygroSensorNo } ::= { wtWebGraphThermHygroBinaryTempValueTable 1 } WtWebGraphThermHygroBinaryTempValueEntry ::= SEQUENCE{ wtWebGraphThermHygroBinaryTempValue INTEGER } wtWebGraphThermHygroBinaryTempValue OBJECT-TYPE SYNTAX INTEGER ACCESS read-only STATUS mandatory DESCRIPTION "Bin�rer Messwert des Sensors in 1/10 Grad Celcius oder 1/10 Prozent, vorzeichenbehaftete Zahl, der Wert 0x4FFF6 zeigt einen ung�ltigen Messwert an." ::= { wtWebGraphThermHygroBinaryTempValueEntry 1 } -- wtWebGraphThermHygroSessCntrl Group -------------------------------- wtWebGraphThermHygroSessCntrlPassword OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Durch Eintrag diese Passworts wird eine Session er�ffnet, die durch Schreiben einer 1 in wtWebGraphThermHygroSessCntrlLogout beendet wird. Gelesen liefert diese Variable einen String der L�nge Null." ::= { wtWebGraphThermHygroSessCntrl 1 } wtWebGraphThermHygroSessCntrlConfigMode OBJECT-TYPE SYNTAX INTEGER { wtWebGraphThermHygroSessCntrl-NoSession(0), wtWebGraphThermHygroSessCntrl-Session(1) } ACCESS read-only STATUS mandatory DESCRIPTION "Wenn sich das Ger�t im SNMP-Konfigurationsmodus befindet, wird der Wert 1 zur�ckgeliefert, ansonsten 0. Ist der ausgelesene Wert 0 obwohl das Passwort geschrieben wurde, befindet sich das Ger�t bereits im Configmodus und die Variablen k�nnen nur gelesen werden." ::= { wtWebGraphThermHygroSessCntrl 2 } wtWebGraphThermHygroSessCntrlLogout OBJECT-TYPE SYNTAX INTEGER ACCESS read-write STATUS mandatory DESCRIPTION "Wenn sich das Ger�t im Konfigurationsmodus befindet und die Session wurde von diesem SNMP Manager er�ffnet, kann durch Schreiben einer 1 in diese Variable die Session beendet werden und alle Parameter werden gespeichert. Durch Schreiben einer 2 werden alle Eingaben wieder verworfen. Gelesen liefert diese Variable den Wert 0." ::= { wtWebGraphThermHygroSessCntrl 3 } wtWebGraphThermHygroSessCntrlAdminPassword OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Dieses Passwort sch�tzt den Zugang der Variablen mit Administratorrechten. Es kann nur vom Administrator ver�ndert werden. Wurde kein Passwort vergeben so ist public das g�ltige Passwort. Gelesen liefert diese Variable einen Null-Pointer." ::= { wtWebGraphThermHygroSessCntrl 4 } wtWebGraphThermHygroSessCntrlConfigPassword OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Dieses Passwort sch�tzt den Zugang der Variablen mit Configuratorrechten. Es kann vom Administrator und vom Configurator ver�ndert werden. Wurde kein Passwort vergeben so ist public das g�ltige Passwort.Gelesen liefert diese Variable einen Null-Pointer." ::= { wtWebGraphThermHygroSessCntrl 5 } -- wtWebGraphThermHygroText Group ------------------------ wtWebGraphThermHygroDeviceName OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Name des Webio's, der userdefinierbar ist." ::= { wtWebGraphThermHygroText 1 } wtWebGraphThermHygroDeviceText OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Beschreibung des Webio's, die userdefinierbar ist." ::= { wtWebGraphThermHygroText 2 } wtWebGraphThermHygroDeviceLocation OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Aufstellort des Webio's, die userdefinierbar ist." ::= { wtWebGraphThermHygroText 3 } wtWebGraphThermHygroDeviceContact OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Kontaktadresse im Servicefall des Webio's, die userdefinierbar ist." ::= { wtWebGraphThermHygroText 4 } -- wtWebGraphThermHygroTimeZone Group ---------------------------- wtWebGraphThermHygroTzOffsetHrs OBJECT-TYPE SYNTAX INTEGER ACCESS read-write STATUS mandatory DESCRIPTION "Offset Stunden der Zeitzone (Winterzeit) in Bezug auf UTC." ::= { wtWebGraphThermHygroTimeZone 1 } wtWebGraphThermHygroTzOffsetMin OBJECT-TYPE SYNTAX INTEGER ACCESS read-write STATUS mandatory DESCRIPTION "Offset Minuten der Zeitzone (Winterzeit) in Bezug auf UTC." ::= { wtWebGraphThermHygroTimeZone 2 } wtWebGraphThermHygroTzEnable OBJECT-TYPE SYNTAX OCTET STRING (SIZE(4)) ACCESS read-write STATUS mandatory DESCRIPTION "Bin�r kodierte Bedingung: Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: ungenutzt Octet 4: Bit 1-7: ungenutzt Bit 0 : Apply Timezone" ::= { wtWebGraphThermHygroTimeZone 3 } wtWebGraphThermHygroStTzOffsetHrs OBJECT-TYPE SYNTAX INTEGER ACCESS read-write STATUS mandatory DESCRIPTION "Offset Stunden der Zeitzone (Sommerzeit) in Bezug auf UTC." ::= { wtWebGraphThermHygroTimeZone 4 } wtWebGraphThermHygroStTzOffsetMin OBJECT-TYPE SYNTAX INTEGER ACCESS read-write STATUS mandatory DESCRIPTION "Offset Minuten der Zeitzone (Sommerzeit) in Bezug auf UTC." ::= { wtWebGraphThermHygroTimeZone 5 } wtWebGraphThermHygroStTzEnable OBJECT-TYPE SYNTAX OCTET STRING (SIZE(4)) ACCESS read-write STATUS mandatory DESCRIPTION "Bin�r kodierte Bedingung: Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: ungenutzt Octet 4: Bit 1-7: ungenutzt Bit 0 : Apply Summertime" ::= { wtWebGraphThermHygroTimeZone 6 } wtWebGraphThermHygroStTzStartMonth OBJECT-TYPE SYNTAX INTEGER { wtWebGraphThermHygroStartMonth-January(1), wtWebGraphThermHygroStartMonth-February(2), wtWebGraphThermHygroStartMonth-March(3), wtWebGraphThermHygroStartMonth-April(4), wtWebGraphThermHygroStartMonth-May(5), wtWebGraphThermHygroStartMonth-June(6), wtWebGraphThermHygroStartMonth-July(7), wtWebGraphThermHygroStartMonth-August(8), wtWebGraphThermHygroStartMonth-September(9), wtWebGraphThermHygroStartMonth-October(10), wtWebGraphThermHygroStartMonth-November(11), wtWebGraphThermHygroStartMonth-December(12) } ACCESS read-write STATUS mandatory DESCRIPTION "Monat, in dem von Winterzeit auf Sommerzeit umgestellt wird." ::= { wtWebGraphThermHygroTimeZone 7 } wtWebGraphThermHygroStTzStartMode OBJECT-TYPE SYNTAX INTEGER { wtWebGraphThermHygroStartMode-first(1), wtWebGraphThermHygroStartMode-second(2), wtWebGraphThermHygroStartMode-third(3), wtWebGraphThermHygroStartMode-fourth(4), wtWebGraphThermHygroStartMode-last(5) } ACCESS read-write STATUS mandatory DESCRIPTION "Modus des Wochentages, bei dem von Winterzeit auf Sommerzeit umgestellt wird." ::= { wtWebGraphThermHygroTimeZone 8 } wtWebGraphThermHygroStTzStartWday OBJECT-TYPE SYNTAX INTEGER { wtWebGraphThermHygroStartWday-Sunday(1), wtWebGraphThermHygroStartWday-Monday(2), wtWebGraphThermHygroStartWday-Tuesday(3), wtWebGraphThermHygroStartWday-Thursday(4), wtWebGraphThermHygroStartWday-Wednesday(5), wtWebGraphThermHygroStartWday-Friday(6), wtWebGraphThermHygroStartWday-Saturday(7) } ACCESS read-write STATUS mandatory DESCRIPTION "Wochentag, bei dem von Winterzeit auf Sommerzeit umgestellt wird" ::= { wtWebGraphThermHygroTimeZone 9 } wtWebGraphThermHygroStTzStartHrs OBJECT-TYPE SYNTAX INTEGER ACCESS read-write STATUS mandatory DESCRIPTION "Stunde, in der die Zeitzone von Winter auf Sommerzeit in Bezug auf UTC umgestellt wird." ::= { wtWebGraphThermHygroTimeZone 10 } wtWebGraphThermHygroStTzStartMin OBJECT-TYPE SYNTAX INTEGER ACCESS read-write STATUS mandatory DESCRIPTION "Minute, in der die Zeitzone von Winter auf Sommerzeit in Bezug auf UTC umgestellt wird." ::= { wtWebGraphThermHygroTimeZone 11 } wtWebGraphThermHygroStTzStopMonth OBJECT-TYPE SYNTAX INTEGER { wtWebGraphThermHygroStopMonth-January(1), wtWebGraphThermHygroStopMonth-February(2), wtWebGraphThermHygroStopMonth-March(3), wtWebGraphThermHygroStopMonth-April(4), wtWebGraphThermHygroStopMonth-May(5), wtWebGraphThermHygroStopMonth-June(6), wtWebGraphThermHygroStopMonth-July(7), wtWebGraphThermHygroStopMonth-August(8), wtWebGraphThermHygroStopMonth-September(9), wtWebGraphThermHygroStopMonth-October(10), wtWebGraphThermHygroStopMonth-November(11), wtWebGraphThermHygroStopMonth-December(12) } ACCESS read-write STATUS mandatory DESCRIPTION "Monat, bei dem von Sommerzeit auf Winterzeit umgestellt wird." ::= { wtWebGraphThermHygroTimeZone 12 } wtWebGraphThermHygroStTzStopMode OBJECT-TYPE SYNTAX INTEGER { wtWebGraphThermHygroStopMode-first(1), wtWebGraphThermHygroStopMode-second(2), wtWebGraphThermHygroStopMode-third(3), wtWebGraphThermHygroStopMode-fourth(4), wtWebGraphThermHygroStopMode-last(5) } ACCESS read-write STATUS mandatory DESCRIPTION "Modus des Wochentages, bei dem von Sommerzeit auf Winterzeit umgestellt wird." ::= { wtWebGraphThermHygroTimeZone 13 } wtWebGraphThermHygroStTzStopWday OBJECT-TYPE SYNTAX INTEGER { wtWebGraphThermHygroStopWday-Sunday(1), wtWebGraphThermHygroStopWday-Monday(2), wtWebGraphThermHygroStopWday-Tuesday(3), wtWebGraphThermHygroStopWday-Thursday(4), wtWebGraphThermHygroStopWday-Wednesday(5), wtWebGraphThermHygroStopWday-Friday(6), wtWebGraphThermHygroStopWday-Saturday(7) } ACCESS read-write STATUS mandatory DESCRIPTION "Wochentage, bei dem von Sommerzeit auf Winterzeit umgestellt wird." ::= { wtWebGraphThermHygroTimeZone 14 } wtWebGraphThermHygroStTzStopHrs OBJECT-TYPE SYNTAX INTEGER ACCESS read-write STATUS mandatory DESCRIPTION "Stunde, in der die Zeitzone von Sommer auf Winterzeit in Bezug auf UTC umgestellt wird." ::= { wtWebGraphThermHygroTimeZone 15 } wtWebGraphThermHygroStTzStopMin OBJECT-TYPE SYNTAX INTEGER ACCESS read-write STATUS mandatory DESCRIPTION "Minute, in der die Zeitzone von Sommer auf Winterzeit in Bezug auf UTC umgestellt wird." ::= { wtWebGraphThermHygroTimeZone 16 } -- wtWebGraphThermHygroTimeServer Group ---------------------------- wtWebGraphThermHygroTimeServer1 OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "IP-Adresse oder Name des 1. Time Servers in ASCII" ::= { wtWebGraphThermHygroTimeServer 1 } wtWebGraphThermHygroTimeServer2 OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "IP-Adresse oder Name des 2. Time Servers in ASCII" ::= { wtWebGraphThermHygroTimeServer 2 } wtWebGraphThermHygroTsEnable OBJECT-TYPE SYNTAX OCTET STRING (SIZE(4)) ACCESS read-write STATUS mandatory DESCRIPTION "Bin�r kodierte Bedingung: Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: ungenutzt Octet 4: Bit 1-7: ungenutzt Bit 0 : Apply Time Server" ::= { wtWebGraphThermHygroTimeServer 3 } wtWebGraphThermHygroTsSyncTime OBJECT-TYPE SYNTAX INTEGER ACCESS read-write STATUS mandatory DESCRIPTION "Die Abfrage der Timeserver erfolgt zu dieser vollen Stunde (0-23)." ::= { wtWebGraphThermHygroTimeServer 4 } -- wtWebGraphThermHygroDeviceClock Group ---------------------------- wtWebGraphThermHygroClockHrs OBJECT-TYPE SYNTAX INTEGER (0..23) ACCESS read-write STATUS mandatory DESCRIPTION "Uhrzeit Stunde, unter Ber�cksichtigung der Zeitzoneneinstellungen." ::= { wtWebGraphThermHygroDeviceClock 1 } wtWebGraphThermHygroClockMin OBJECT-TYPE SYNTAX INTEGER (0..59) ACCESS read-write STATUS mandatory DESCRIPTION "Uhrzeit Minute, unter Ber�cksichtigung der Zeitzoneneinstellungen." ::= { wtWebGraphThermHygroDeviceClock 2 } wtWebGraphThermHygroClockDay OBJECT-TYPE SYNTAX INTEGER (1..31) ACCESS read-write STATUS mandatory DESCRIPTION "Datum: Tag des Monats, unter Ber�cksichtigung der Zeitzoneneinstellungen." ::= { wtWebGraphThermHygroDeviceClock 3 } wtWebGraphThermHygroClockMonth OBJECT-TYPE SYNTAX INTEGER { wtWebGraphThermHygroClockMonth-January(1), wtWebGraphThermHygroClockMonth-February(2), wtWebGraphThermHygroClockMonth-March(3), wtWebGraphThermHygroClockMonth-April(4), wtWebGraphThermHygroClockMonth-May(5), wtWebGraphThermHygroClockMonth-June(6), wtWebGraphThermHygroClockMonth-July(7), wtWebGraphThermHygroClockMonth-August(8), wtWebGraphThermHygroClockMonth-September(9), wtWebGraphThermHygroClockMonth-October(10), wtWebGraphThermHygroClockMonth-November(11), wtWebGraphThermHygroClockMonth-December(12) } ACCESS read-write STATUS mandatory DESCRIPTION "Datum: Monat, unter Ber�cksichtigung der Zeitzoneneinstellungen." ::= { wtWebGraphThermHygroDeviceClock 4 } wtWebGraphThermHygroClockYear OBJECT-TYPE SYNTAX INTEGER (0..65535) ACCESS read-write STATUS mandatory DESCRIPTION "Datum: Jahr, unter Ber�cksichtigung der Zeitzoneneinstellungen." ::= { wtWebGraphThermHygroDeviceClock 5 } -- wtWebGraphThermHygroNetwork Group ------------------------ wtWebGraphThermHygroIpAddress OBJECT-TYPE SYNTAX IpAddress ACCESS read-write STATUS mandatory DESCRIPTION "IP-Adresse des Webio's" ::= { wtWebGraphThermHygroNetwork 1 } wtWebGraphThermHygroSubnetMask OBJECT-TYPE SYNTAX IpAddress ACCESS read-write STATUS mandatory DESCRIPTION "Subnetmask des Webio's" ::= { wtWebGraphThermHygroNetwork 2 } wtWebGraphThermHygroGateway OBJECT-TYPE SYNTAX IpAddress ACCESS read-write STATUS mandatory DESCRIPTION "Gateway des Webio's" ::= { wtWebGraphThermHygroNetwork 3 } wtWebGraphThermHygroDnsServer1 OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "IP-Adresse des 1. DNS Servers in ASCII" ::= { wtWebGraphThermHygroNetwork 4 } wtWebGraphThermHygroDnsServer2 OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "IP-Adresse des 2. DNS Servers in ASCII" ::= { wtWebGraphThermHygroNetwork 5 } wtWebGraphThermHygroAddConfig OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Zusatzkonfiguration (1=EIN): Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: ungenutzt Octet 4: Bit 0: BOOTP disable " ::= { wtWebGraphThermHygroNetwork 6 } -- wtWebGraphThermHygroHTTP Group -------------------------------- wtWebGraphThermHygroStartup OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Legt die Startseite fest, nur eine Auswahl zul�ssig: Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: ungenutzt Octet 4: Bit 2: user.htm Bit 1: home.htm Bit 0: index.htm" ::= { wtWebGraphThermHygroHTTP 1 } wtWebGraphThermHygroGetHeaderEnable OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Wenn freigeschaltet, wird eine Erkennungssequenz nach jeder GET-Anfrage der Antwort vorangestellt: Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: ungenutzt Octet 4: Bit 1-7: ungenutzt Bit 0 : GET Header enable" ::= { wtWebGraphThermHygroHTTP 2 } wtWebGraphThermHygroHttpPort OBJECT-TYPE SYNTAX INTEGER (1..65535) ACCESS read-write STATUS mandatory DESCRIPTION "HTTP Port, an den die TCP-Anfragen gesendet werden k�nnen (Default: 80)." ::= { wtWebGraphThermHygroHTTP 3 } -- wtWebGraphThermHygroMail Group -------------------------------- wtWebGraphThermHygroMailAdName OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Ger�tename f�r die Mail-Administration." ::= { wtWebGraphThermHygroMail 1 } wtWebGraphThermHygroMailReply OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Reply-Adresse f�r den Mailserver." ::= { wtWebGraphThermHygroMail 2 } wtWebGraphThermHygroMailServer OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Name oder Ip-Adresse des Mailservers." ::= { wtWebGraphThermHygroMail 3 } wtWebGraphThermHygroMailEnable OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Erlaubt das versenden von Mails: Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: ungenutzt Octet 4: Bit 1-7: ungenutzt Bit 0 : Mail enable" ::= { wtWebGraphThermHygroMail 4 } wtWebGraphThermHygroMailAuthentication OBJECT-TYPE SYNTAX OCTET STRING (SIZE (4)) ACCESS read-write STATUS mandatory DESCRIPTION "Auswahl des Authenticationmodi: die OCTETS werden bitweise interpretiert, es darf nur ein Bit gesetzt sein. Bit 0 entspricht dem LSB. OCTET 1: ungenutzt OCTET 2: ungenutzt OCTET 3: ungenutzt OCTET 4: Bit 3-7: ungenutzt Bit 2 : POP3 vor SMTP Bit 1 : ESMTP, Authentifizierung Bit 0 : SMTP, keine Authentifizierung" ::= { wtWebGraphThermHygroMail 5 } wtWebGraphThermHygroMailAuthUser OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Name des Postfaches / Postfachusers." ::= { wtWebGraphThermHygroMail 6 } wtWebGraphThermHygroMailAuthPassword OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Passwort zur Identifizierung des Users." ::= { wtWebGraphThermHygroMail 7 } wtWebGraphThermHygroMailPop3Server OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Name oder Ip-Adresse des POP3 Servers." ::= { wtWebGraphThermHygroMail 8 } -- wtWebGraphThermHygroSNMP Group -------------------------------- wtWebGraphThermHygroSnmpCommunityStringRead OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Community string read only" ::= { wtWebGraphThermHygroSNMP 1 } wtWebGraphThermHygroSnmpCommunityStringReadWrite OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Community string read / write" ::= { wtWebGraphThermHygroSNMP 2 } wtWebGraphThermHygroSystemTrapManagerIP OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Name oder IP-Adresse des SNMP-Managers, der den System Trap entgegen nimmt." ::= { wtWebGraphThermHygroSNMP 3 } wtWebGraphThermHygroSystemTrapEnable OBJECT-TYPE SYNTAX OCTET STRING (SIZE(4)) ACCESS read-write STATUS mandatory DESCRIPTION "Bin�r kodierte Triggerbedingungen: Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: ungenutzt Octet 4: Bit 0: ColdStart Bit 1: WarmStart Bit 2-7: ungenutzt" ::= { wtWebGraphThermHygroSNMP 4 } wtWebGraphThermHygroSnmpEnable OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Bit 0 schaltet SNMP frei: Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: ungenutzt Octet 4: Bit 1-7: ungenutzt Bit 0 : SNMP enable" ::= { wtWebGraphThermHygroSNMP 5 } wtWebGraphThermHygroSnmpCommunityStringTrap OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Community string der Traps." ::= { wtWebGraphThermHygroSNMP 6 } -- wtWebGraphThermHygroUDP Group -------------------------------- wtWebGraphThermHygroUdpPort OBJECT-TYPE SYNTAX INTEGER (1..65535) ACCESS read-write STATUS mandatory DESCRIPTION "Udp Port, an den die Udp-Anfragen GET Single gesendet werden k�nnen" ::= { wtWebGraphThermHygroUDP 1 } wtWebGraphThermHygroUdpEnable OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Schaltet UDP frei: Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: ungenutzt Octet 4: Bit 1-7: ungenutzt Bit 0 : UDP enable" ::= { wtWebGraphThermHygroUDP 2 } -- wtWebGraphThermHygroSyslog Group -------------------------------- wtWebGraphThermHygroSyslogServerIP OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Name oder IP-Adresse des Syslog-Servers, der die Syslog-Nachricht entgegen nimmt." ::= { wtWebGraphThermHygroSyslog 1 } wtWebGraphThermHygroSyslogServerPort OBJECT-TYPE SYNTAX INTEGER ACCESS read-write STATUS mandatory DESCRIPTION "Portnummer des Syslog-Servers, default ist 514." ::= { wtWebGraphThermHygroSyslog 2 } wtWebGraphThermHygroSyslogSystemMessagesEnable OBJECT-TYPE SYNTAX OCTET STRING (SIZE(4)) ACCESS read-write STATUS mandatory DESCRIPTION "Bin�r kodierte Triggerbedingungen: Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: ungenutzt Octet 4: Bit 0: ColdStart Bit 1: WarmStart Bit 2: Diag Messages Bit 3-7: ungenutzt" ::= { wtWebGraphThermHygroSyslog 3 } wtWebGraphThermHygroSyslogEnable OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Bit 0 schaltet Syslog frei: Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: ungenutzt Octet 4: Bit 1-7: ungenutzt Bit 0 : Syslog enable" ::= { wtWebGraphThermHygroSyslog 4 } -- wtWebGraphThermHygroFTP Group -------------------------------- wtWebGraphThermHygroFTPServerIP OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Name oder IP-Adresse des FTP-Servers, der die FTP-Nachricht entgegen nimmt." ::= { wtWebGraphThermHygroFTP 1 } wtWebGraphThermHygroFTPServerControlPort OBJECT-TYPE SYNTAX INTEGER ACCESS read-write STATUS mandatory DESCRIPTION "Portnummer des Kontrollports des FTP-Servers, default ist 21." ::= { wtWebGraphThermHygroFTP 2 } wtWebGraphThermHygroFTPUserName OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "FTP Loggin: Name des Anwenders." ::= { wtWebGraphThermHygroFTP 3 } wtWebGraphThermHygroFTPPassword OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "FTP Loggin: Passwort des Anwenders." ::= { wtWebGraphThermHygroFTP 4 } wtWebGraphThermHygroFTPAccount OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "FTP Loggin: Account-Name des Anwenderkontos." ::= { wtWebGraphThermHygroFTP 5 } wtWebGraphThermHygroFTPOption OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Bit 0 schaltet den FTP Server in den Passiv-Modus, der Client kontrolliert die Datenverbindung allein: Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: ungenutzt Octet 4: Bit 1-7: ungenutzt Bit 0 : PASV enable" ::= { wtWebGraphThermHygroFTP 6 } wtWebGraphThermHygroFTPEnable OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Bit 0 schaltet FTP frei: Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: ungenutzt Octet 4: Bit 1-7: ungenutzt Bit 0 : FTP enable" ::= { wtWebGraphThermHygroFTP 7 } -- wtWebGraphThermHygroDatalogger Group -------------------------------- wtWebGraphThermHygroLoggerTimebase OBJECT-TYPE SYNTAX INTEGER { wtWebGraphThermHygroDatalogger-1Min(1), wtWebGraphThermHygroDatalogger-5Min(2), wtWebGraphThermHygroDatalogger-15Min(3), wtWebGraphThermHygroDatalogger-60Min(4) } ACCESS read-write STATUS mandatory DESCRIPTION "Zeitbasis des Dataloggers" ::= { wtWebGraphThermHygroDatalogger 1 } wtWebGraphThermHygroLoggerSensorSel OBJECT-TYPE SYNTAX OCTET STRING (SIZE(4)) ACCESS read-write STATUS mandatory DESCRIPTION "Bin�r kodierte Triggerbedingungen: Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: ungenutzt Octet 4: Bit 0 : Sensor 1 Bit 1 : Sensor 2 Bit 2-7: ungenutzt" ::= { wtWebGraphThermHygroDatalogger 2 } wtWebGraphThermHygroDisplaySensorSel OBJECT-TYPE SYNTAX OCTET STRING (SIZE(4)) ACCESS read-write STATUS mandatory DESCRIPTION "Die hier ausgew�hlten Sensoren werden grafisch angezeigt: Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: ungenutzt Octet 4: Bit 0: Sensor 1 Bit 1: Sensor 2 Bit 2-7: ungenutzt" ::= { wtWebGraphThermHygroDatalogger 3 } -- Graph Sensor Color Table wtWebGraphThermHygroSensorColorTable OBJECT-TYPE SYNTAX SEQUENCE OF WtWebGraphThermHygroSensorColorEntry ACCESS not-accessible STATUS mandatory DESCRIPTION "Liste der Sensorfarben in der grafischen Darstellung." ::= { wtWebGraphThermHygroDatalogger 4 } wtWebGraphThermHygroSensorColorEntry OBJECT-TYPE SYNTAX WtWebGraphThermHygroSensorColorEntry ACCESS not-accessible STATUS mandatory DESCRIPTION "Nummern der Sensoren." INDEX { wtWebGraphThermHygroSensorNo } ::= { wtWebGraphThermHygroSensorColorTable 1 } WtWebGraphThermHygroSensorColorEntry ::= SEQUENCE { wtWebGraphThermHygroSensorColor OCTET STRING } wtWebGraphThermHygroSensorColor OBJECT-TYPE SYNTAX OCTET STRING (SIZE(3)) ACCESS read-write STATUS mandatory DESCRIPTION "Farbe des Sensors in der grafischen Darstellung." ::= { wtWebGraphThermHygroSensorColorEntry 1 } wtWebGraphThermHygroAutoScaleEnable OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Bit 0 schaltet Vertical Auto Scale frei: Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: ungenutzt Octet 4: Bit 1-7: ungenutzt Bit 0 : Auto Scale enable" ::= { wtWebGraphThermHygroDatalogger 5 } wtWebGraphThermHygroVerticalUpperLimit OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Definiert die maximal sichtbare Temperatur f�r die grafische Darstellung. Temperatur in Grad mit maximal 2 Nachkommastellen." ::= { wtWebGraphThermHygroDatalogger 6 } wtWebGraphThermHygroVerticalLowerLimit OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Definiert die Minal sichtbare Temperatur f�r die grafische Darstellung. Temperatur in Grad mit maximal 2 Nachkommastellen." ::= { wtWebGraphThermHygroDatalogger 7 } wtWebGraphThermHygroHorizontalZoom OBJECT-TYPE SYNTAX INTEGER { wtWebGraphThermHygroZoom-25Min(1), wtWebGraphThermHygroZoom-75Min(2), wtWebGraphThermHygroZoom-5Std(3), wtWebGraphThermHygroZoom-30Std(4), wtWebGraphThermHygroZoom-5Tage(5), wtWebGraphThermHygroZoom-25Tage(6) } ACCESS read-write STATUS mandatory DESCRIPTION "Definiert den default Zoom." ::= { wtWebGraphThermHygroDatalogger 8 } -- wtWebGraphThermHygroAlarm Group ---------------------------- wtWebGraphThermHygroAlarmCount OBJECT-TYPE SYNTAX INTEGER (1..8) ACCESS read-only STATUS mandatory DESCRIPTION "Anzahl der unabh�ngigen Alarme." ::= { wtWebGraphThermHygroAlarm 1 } -- Alarm Interface Table wtWebGraphThermHygroAlarmIfTable OBJECT-TYPE SYNTAX SEQUENCE OF WtWebGraphThermHygroAlarmIfEntry ACCESS not-accessible STATUS mandatory DESCRIPTION "Liste der Alarmindizes." ::= { wtWebGraphThermHygroAlarm 2 } wtWebGraphThermHygroAlarmIfEntry OBJECT-TYPE SYNTAX WtWebGraphThermHygroAlarmIfEntry ACCESS not-accessible STATUS mandatory DESCRIPTION "Nummern der Alarme." INDEX { wtWebGraphThermHygroAlarmNo } ::= { wtWebGraphThermHygroAlarmIfTable 1 } WtWebGraphThermHygroAlarmIfEntry ::= SEQUENCE { wtWebGraphThermHygroAlarmNo INTEGER } wtWebGraphThermHygroAlarmNo OBJECT-TYPE SYNTAX INTEGER (1..8) ACCESS read-only STATUS mandatory DESCRIPTION "Nummer des Alarms." ::= { wtWebGraphThermHygroAlarmIfEntry 1 } -- Alarm Table wtWebGraphThermHygroAlarmTable OBJECT-TYPE SYNTAX SEQUENCE OF WtWebGraphThermHygroAlarmEntry ACCESS not-accessible STATUS mandatory DESCRIPTION "Liste der Einstellungen abh�ngig von der Alarmnummer." ::= { wtWebGraphThermHygroAlarm 3 } wtWebGraphThermHygroAlarmEntry OBJECT-TYPE SYNTAX WtWebGraphThermHygroAlarmEntry ACCESS not-accessible STATUS mandatory DESCRIPTION "Alarm-Parameter der einzelnen Alarme." INDEX { wtWebGraphThermHygroAlarmNo } ::= { wtWebGraphThermHygroAlarmTable 1 } WtWebGraphThermHygroAlarmEntry ::= SEQUENCE { wtWebGraphThermHygroAlarmTrigger OCTET STRING, wtWebGraphThermHygroAlarmMin OCTET STRING, wtWebGraphThermHygroAlarmMax OCTET STRING, wtWebGraphThermHygroAlarmHysteresis OCTET STRING, wtWebGraphThermHygroAlarmDelay OCTET STRING, wtWebGraphThermHygroAlarmInterval OCTET STRING, wtWebGraphThermHygroAlarmEnable OCTET STRING, wtWebGraphThermHygroAlarmEMailAddr OCTET STRING, wtWebGraphThermHygroAlarmMailSubject OCTET STRING, wtWebGraphThermHygroAlarmMailText OCTET STRING, wtWebGraphThermHygroAlarmManagerIP OCTET STRING, wtWebGraphThermHygroAlarmTrapText OCTET STRING, wtWebGraphThermHygroAlarmMailOptions OCTET STRING, wtWebGraphThermHygroAlarmTcpIpAddr OCTET STRING, wtWebGraphThermHygroAlarmTcpPort INTEGER, wtWebGraphThermHygroAlarmTcpText OCTET STRING, wtWebGraphThermHygroAlarmClearMailSubject OCTET STRING, wtWebGraphThermHygroAlarmClearMailText OCTET STRING, wtWebGraphThermHygroAlarmClearTrapText OCTET STRING, wtWebGraphThermHygroAlarmClearTcpText OCTET STRING, wtWebGraphThermHygroAlarmDeltaTemp OCTET STRING, wtWebGraphThermHygroAlarmRHMin OCTET STRING, wtWebGraphThermHygroAlarmRHMax OCTET STRING, wtWebGraphThermHygroAlarmRHHysteresis OCTET STRING, wtWebGraphThermHygroAlarmAHMin OCTET STRING, wtWebGraphThermHygroAlarmAHMax OCTET STRING, wtWebGraphThermHygroAlarmSyslogIpAddr OCTET STRING, wtWebGraphThermHygroAlarmSyslogPort INTEGER, wtWebGraphThermHygroAlarmSyslogText OCTET STRING, wtWebGraphThermHygroAlarmSyslogClearText OCTET STRING, wtWebGraphThermHygroAlarmFtpDataPort OCTET STRING, wtWebGraphThermHygroAlarmFtpFileName OCTET STRING, wtWebGraphThermHygroAlarmFtpText OCTET STRING, wtWebGraphThermHygroAlarmFtpClearText OCTET STRING, wtWebGraphThermHygroAlarmFtpOption OCTET STRING, wtWebGraphThermHygroAlarmTimerCron OCTET STRING } wtWebGraphThermHygroAlarmTrigger OBJECT-TYPE SYNTAX OCTET STRING (SIZE(4)) ACCESS read-write STATUS mandatory DESCRIPTION "Bin�r kodierte Triggerbedingungen: Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: Bit 0: Timer Bit 1: ColdStart Bit 2: WarmStart Bit 3: Sensor lost Bit 4-7 ungenutzt Octet 4: Bit 0: Sensor 1 Bit 1: Sensor 2 Bit 2-7: ungenutzt" ::= { wtWebGraphThermHygroAlarmEntry 1 } wtWebGraphThermHygroAlarmMin OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Minimum Temperatur in Grad mit maximal 2 Nachkommastellen." ::= { wtWebGraphThermHygroAlarmEntry 2 } wtWebGraphThermHygroAlarmMax OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Maximum Temperatur in Grad mit maximal 2 Nachkommastellen." ::= { wtWebGraphThermHygroAlarmEntry 3 } wtWebGraphThermHygroAlarmHysteresis OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Bei �berschreiten der Temperatur Min+Hyst. und Unterschreiten der Temperatur Max-Hyst. wird der Alarm gel�scht: pos. Temperaturdifferenz in Grad mit maximal 2 Nachkommastellen." ::= { wtWebGraphThermHygroAlarmEntry 4 } wtWebGraphThermHygroAlarmDelay OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Der Alarm wird erst ausgel�st, wenn f�r die Dauer dieser Zeit die Alarmbedingungen erf�llt sind (Zeit in Minuten)." ::= { wtWebGraphThermHygroAlarmEntry 5 } wtWebGraphThermHygroAlarmInterval OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Sendeintervall von Alarmmeldungen." ::= { wtWebGraphThermHygroAlarmEntry 6 } wtWebGraphThermHygroAlarmEnable OBJECT-TYPE SYNTAX OCTET STRING (SIZE(4)) ACCESS read-write STATUS mandatory DESCRIPTION "Bin�r kodierte Alarmfreischaltung: Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: ungenutzt Octet 4: Bit 0: Mail Enable Bit 1: SNMP Trap enable Bit 2: TCP Client enable" ::= { wtWebGraphThermHygroAlarmEntry 7 } wtWebGraphThermHygroAlarmEMailAddr OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "E-Mail Adresse, an die die Alarmmail gesendet werden soll." ::= { wtWebGraphThermHygroAlarmEntry 8 } wtWebGraphThermHygroAlarmMailSubject OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Betreffzeile der Alarmmail." ::= { wtWebGraphThermHygroAlarmEntry 9 } wtWebGraphThermHygroAlarmMailText OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Text der Alarmmail: <Tn> liefert die aktuelle Temperatur; n = 1..8. <Z> liefert die aktuelle Zeit." ::= { wtWebGraphThermHygroAlarmEntry 10 } wtWebGraphThermHygroAlarmManagerIP OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Name oder IP-Adresse des SNMP-Managers, der den Trap entgegen nimmt." ::= { wtWebGraphThermHygroAlarmEntry 11 } wtWebGraphThermHygroAlarmTrapText OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Dieser Text wird im Trapbody gesendet." ::= { wtWebGraphThermHygroAlarmEntry 12 } wtWebGraphThermHygroAlarmMailOptions OBJECT-TYPE SYNTAX OCTET STRING (SIZE(4)) ACCESS read-write STATUS mandatory DESCRIPTION "Bin�r kodierte Optionen: Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: ungenutzt Octet 4: Bit 0 : Attach Thermo.csv Enable Bit 1 : CSV-Data_since_last_report Bit 2-7: ungenutzt" ::= { wtWebGraphThermHygroAlarmEntry 13 } wtWebGraphThermHygroAlarmTcpIpAddr OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "IP-Adresse des Ger�tes, dass eine TCP-Nachricht empfangen soll." ::= { wtWebGraphThermHygroAlarmEntry 14 } wtWebGraphThermHygroAlarmTcpPort OBJECT-TYPE SYNTAX INTEGER (1..65535) ACCESS read-write STATUS mandatory DESCRIPTION "Portnummer des Ger�tes, dass eine TCP-Nachricht empfangen soll." ::= { wtWebGraphThermHygroAlarmEntry 15 } wtWebGraphThermHygroAlarmTcpText OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Dieser Text wird im TCP Trapbody gesendet." ::= { wtWebGraphThermHygroAlarmEntry 16 } wtWebGraphThermHygroAlarmClearMailSubject OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Betreffzeile der Alarm Clear Mail." ::= { wtWebGraphThermHygroAlarmEntry 17 } wtWebGraphThermHygroAlarmClearMailText OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Text der Alarm Clear Mail: <Tn> liefert die aktuelle Temperatur; n = 1..8. <Z> liefert die aktuelle Zeit." ::= { wtWebGraphThermHygroAlarmEntry 18 } wtWebGraphThermHygroAlarmClearTrapText OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Dieser Text wird im Trapbody gesendet." ::= { wtWebGraphThermHygroAlarmEntry 19 } wtWebGraphThermHygroAlarmClearTcpText OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Dieser Text wird im TCP Trapbody gesendet." ::= { wtWebGraphThermHygroAlarmEntry 20 } wtWebGraphThermHygroAlarmDeltaTemp OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Maximale Temperatur�nderung in Grad/min mit maximal 2 Nachkommastellen. Es wird ein Mittelwert einer Periode von 5 Minuten verglichen." ::= { wtWebGraphThermHygroAlarmEntry 21 } wtWebGraphThermHygroAlarmRHMin OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Minimum rel. Feuchte in Prozent mit maximal 2 Nachkommastellen." ::= { wtWebGraphThermHygroAlarmEntry 22 } wtWebGraphThermHygroAlarmRHMax OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Maximum rel.Feuchte in Prozent mit maximal 2 Nachkommastellen." ::= { wtWebGraphThermHygroAlarmEntry 23 } wtWebGraphThermHygroAlarmRHHysteresis OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Hysterese rel. Feuchte in Prozent mit maximal 2 Nachkommastellen." ::= { wtWebGraphThermHygroAlarmEntry 24 } wtWebGraphThermHygroAlarmAHMin OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Minimum abs. Feuchte in g/(m*m*m) mit maximal 2 Nachkommastellen." ::= { wtWebGraphThermHygroAlarmEntry 25 } wtWebGraphThermHygroAlarmAHMax OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Maximum abs. Feuchte in g/(m*m*m) mit maximal 2 Nachkommastellen." ::= { wtWebGraphThermHygroAlarmEntry 26 } wtWebGraphThermHygroAlarmSyslogIpAddr OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "IP-Adresse des Ger�tes, dass eine Syslog-Nachricht empfangen soll." ::= { wtWebGraphThermHygroAlarmEntry 27 } wtWebGraphThermHygroAlarmSyslogPort OBJECT-TYPE SYNTAX INTEGER (1..65535) ACCESS read-write STATUS mandatory DESCRIPTION "Portnummer des Ger�tes, dass eine Syslog-Nachricht empfangen soll, default Port ist 514." ::= { wtWebGraphThermHygroAlarmEntry 28 } wtWebGraphThermHygroAlarmSyslogText OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Jede Syslog-Nachricht beginnt mit einem Priorit�tsschl�ssel, einem Zeitstempel und der Host Ip-Nummer gem�� RFC3164. Dieser Text wird im darauf folgenden Syslog-Nachrichtenteil gesendet." ::= { wtWebGraphThermHygroAlarmEntry 29 } wtWebGraphThermHygroAlarmSyslogClearText OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Jede Syslog-Nachricht beginnt mit einem Priorit�tsschl�ssel, einem Zeitstempel und der Host Ip-Nummer gem�� RFC3164. Dieser Text wird im darauf folgenden Syslog-Nachrichtenteil gesendet." ::= { wtWebGraphThermHygroAlarmEntry 30 } wtWebGraphThermHygroAlarmFtpDataPort OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Lokaler Datenport: 1..65535 oder AUTO: der n�chste ungenutzte Port wird automatisch zugewiesen." ::= { wtWebGraphThermHygroAlarmEntry 31 } wtWebGraphThermHygroAlarmFtpFileName OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Der File Name kann auch Pfadinformationen enthalten." ::= { wtWebGraphThermHygroAlarmEntry 32 } wtWebGraphThermHygroAlarmFtpText OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Dieser Text wird in die angegebene Datei geschrieben." ::= { wtWebGraphThermHygroAlarmEntry 33 } wtWebGraphThermHygroAlarmFtpClearText OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Dieser Text wird in die angegebene Datei geschrieben." ::= { wtWebGraphThermHygroAlarmEntry 34 } wtWebGraphThermHygroAlarmFtpOptions OBJECT-TYPE SYNTAX OCTET STRING (SIZE(4)) ACCESS read-write STATUS mandatory DESCRIPTION "Bin�r kodierte Optionen, nur eine Option darf eingeschaltet sein: STORE = Datei ersetzen APPEND = Nachrichten ans Dateiende anh�ngen Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: ungenutzt Octet 4: Bit 0 : STORE Bit 1 : APPEND Bit 2-7: ungenutzt" ::= { wtWebGraphThermHygroAlarmEntry 35 } wtWebGraphThermHygroAlarmTimerCron OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Diser Eintrag enth�lt 5 Leerzeichen- oder Tab-getrente Felder zur Definition von zeitgesteuerten Ausgabe von Alarmnachrichten. Jedes der 5 Felder muss mit der aktuelle Zeit �bereinstimmen, um das Senden einer Nachricht auszul�sen, Wochentag und Monatstag werden getrennt ausgewertet. Feld m�glicher Zahlenbereich 1: Minuten 0-59 2. Stunden 0-23, 0 ist Mitternacht 3: Monatstag 1-31 4: Monat 1-12 5: Wochentag 0-6, 0 ist Sonntag F�r jedes Feld kann nun angegeben werden: - eine ganze Zahl aus dem angegebenen Zahlenbereich - ein Teil-Zahlenbereich (wie z.B. 1-5) innerhalb des angegebenen Zahlenbereichs - eine mit Komma getrennte Liste von ganzen Zahlen oder Tei-Zahlenbereichen (wie z.B. 2,4-6,8,10) - ein Stern * (deckt den gesamten erlaubten Bereich ab) - ein / gibt die Schrittweite bei Teil-Zahlenbereichen an (z.B. 0-59/2 jede zweite Minute)." ::= { wtWebGraphThermHygroAlarmEntry 36 } -- wtWebGraphThermHygroPorts Group ---------------------------- -- Table wtWebGraphThermHygroPortTable OBJECT-TYPE SYNTAX SEQUENCE OF WtWebGraphThermHygroPortEntry ACCESS not-accessible STATUS mandatory DESCRIPTION "Liste der Einstellungen abh�ngig von der Sensornummer." ::= { wtWebGraphThermHygroPorts 1 } wtWebGraphThermHygroPortEntry OBJECT-TYPE SYNTAX WtWebGraphThermHygroPortEntry ACCESS not-accessible STATUS mandatory DESCRIPTION "Sensor-Parameter der einzelnen Sensoren." INDEX { wtWebGraphThermHygroSensorNo } ::= { wtWebGraphThermHygroPortTable 1 } WtWebGraphThermHygroPortEntry ::= SEQUENCE { wtWebGraphThermHygroPortName OCTET STRING, wtWebGraphThermHygroPortText OCTET STRING, wtWebGraphThermHygroPortOffset1 OCTET STRING, wtWebGraphThermHygroPortTemperature1 OCTET STRING, wtWebGraphThermHygroPortOffset2 OCTET STRING, wtWebGraphThermHygroPortTemperature2 OCTET STRING, wtWebGraphThermHygroPortComment OCTET STRING, wtWebGraphThermHygroPortSensorSelect OCTET STRING } wtWebGraphThermHygroPortName OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Name des Sensors." ::= { wtWebGraphThermHygroPortEntry 1 } wtWebGraphThermHygroPortText OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Beschreibung des Sensors." ::= { wtWebGraphThermHygroPortEntry 2 } wtWebGraphThermHygroPortOffset1 OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Offset-Kalibrierung: Offset in Grad mit max. 2 Nachkommastellen bei Temperatur 1." ::= { wtWebGraphThermHygroPortEntry 3 } wtWebGraphThermHygroPortTemperature1 OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Offset-Kalibrierung: Referenztemperatur 1 in Grad mit max. 2 Nachkommastellen." ::= { wtWebGraphThermHygroPortEntry 4 } wtWebGraphThermHygroPortOffset2 OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Offset-Kalibrierung: Offset in Grad mit max. 2 Nachkommastellen bei Temperatur 2." ::= { wtWebGraphThermHygroPortEntry 5 } wtWebGraphThermHygroPortTemperature2 OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Offset-Kalibrierung: Referenztemperatur 2 in Grad mit max. 2 Nachkommastellen." ::= { wtWebGraphThermHygroPortEntry 6 } wtWebGraphThermHygroPortComment OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Offset-Kalibrierung: Angaben zu Datum, Referenzger�t, Operator." ::= { wtWebGraphThermHygroPortEntry 7 } wtWebGraphThermHygroPortSensorSelect OBJECT-TYPE SYNTAX OCTET STRING (SIZE(4)) ACCESS read-write STATUS mandatory DESCRIPTION "Bin�r kodierte Optionen f�r Sensor 1: Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: ungenutzt Octet 4: Bit 0 : PT100 Bit 1 : PT1000 (default) Bit 2-7: ungenutzt Bin�r kodierte Optionen f�r Sensor 2: Octet 1: ungenutzt Octet 2: ungenutzt Octet 3: ungenutzt Octet 4: Bit 0 : W&T Sensor rel. Feuchte (default) Bit 1 : Skalar 0-2.5V Bit 2 : Disconnect Bit 3-7: ungenutzt" ::= { wtWebGraphThermHygroPortEntry 8 } -- wtWebGraphThermHygroManufact Group ---------------------------- wtWebGraphThermHygroMfName OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Hersteller Firmenname" ::= { wtWebGraphThermHygroManufact 1 } wtWebGraphThermHygroMfAddr OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Adresse des Herstellers." ::= { wtWebGraphThermHygroManufact 2 } wtWebGraphThermHygroMfHotline OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Hotline des Herstellers." ::= { wtWebGraphThermHygroManufact 3 } wtWebGraphThermHygroMfInternet OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Internet-Adresse des Herstellers." ::= { wtWebGraphThermHygroManufact 4 } wtWebGraphThermHygroMfDeviceTyp OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Ger�tebezeichnung (Typ) des Herstellers." ::= { wtWebGraphThermHygroManufact 5 } wtWebGraphThermHygroMfOrderNo OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-write STATUS mandatory DESCRIPTION "Bestellnummer des Ger�tes des Herstellers." ::= { wtWebGraphThermHygroManufact 6 } -- wtWebGraphThermHygroDiag Group ---------------------------- wtWebGraphThermHygroDiagErrorCount OBJECT-TYPE SYNTAX INTEGER ACCESS read-only STATUS mandatory DESCRIPTION "Anzahl der Eintr�ge im Fehlerspeicher." ::= { wtWebGraphThermHygroDiag 1 } wtWebGraphThermHygroDiagBinaryError OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-only STATUS mandatory DESCRIPTION "Bitweise kodierte Fehlerursachen." ::= { wtWebGraphThermHygroDiag 2 } wtWebGraphThermHygroDiagErrorIndex OBJECT-TYPE SYNTAX INTEGER ACCESS read-write STATUS mandatory DESCRIPTION "Der Wert dieser Variable wird als Index f�r die Ausgabe eines Fehlertextes (wtWebGraphThermHygroDiagErrorMessage) verwendet. Die Fehler im Speicher sind von 1 bis wtWebGraphThermHygroDiagErrorCount durchnummeriert. Nach dem Schreiben einens g�ltigen Wertes in diese Variable kann der dazugeh�rige Text in wtWebGraphThermHygroDiagErrorMessage ausgelesen werden." ::= { wtWebGraphThermHygroDiag 3 } wtWebGraphThermHygroDiagErrorMessage OBJECT-TYPE SYNTAX OCTET STRING ACCESS read-only STATUS mandatory DESCRIPTION "Abh�ngig vom Wert wtWebGraphThermHygroDiagErrorIndex wird ein Fehlertext angezeigt." ::= { wtWebGraphThermHygroDiag 4 } wtWebGraphThermHygroDiagErrorClear OBJECT-TYPE SYNTAX INTEGER ACCESS write-only STATUS mandatory DESCRIPTION "Durch Schreiben eine 1 in diese Variable, wird der Fehlerspeicher gel�scht." ::= { wtWebGraphThermHygroDiag 5 } -- TRAP DEFINITIONS -- Diese SNMP-Traps werden gesendet, wenn die Alarmtriggerbedingungen zutreffen wtWebGraphThermHygroAlert1 TRAP-TYPE ENTERPRISE wtWebGraphThermHygro VARIABLES { wtWebGraphThermHygroAlarmTrapText } DESCRIPTION "Dieser SNMP-Trap wird generiert, wenn der Alarm 1 ausgel�st wurde, das SNMP-Trap-Enable gesetzt ist und SNMP in Enable freigegeben ist." -- NMS trap annotation --#TYPE "Thermo-Hygrometer Alarm 1" --#SUMMARY "Alarm 1: %s" --#ARGUMENTS {0} --#SEVERITY INFORMATIONAL --#TIMEINDEX 99 --#STATE OPERATIONAL ::= 31 wtWebGraphThermHygroAlert2 TRAP-TYPE ENTERPRISE wtWebGraphThermHygro VARIABLES { wtWebGraphThermHygroAlarmTrapText } DESCRIPTION "Dieser SNMP-Trap wird generiert, wenn der Alarm 2 ausgel�st wurde, das SNMP-Trap-Enable gesetzt ist und SNMP in Enable freigegeben ist." -- NMS trap annotation --#TYPE "Thermo-Hygrometer Alarm 2" --#SUMMARY "Alarm 2: %s" --#ARGUMENTS {0} --#SEVERITY INFORMATIONAL --#TIMEINDEX 99 --#STATE OPERATIONAL ::= 32 wtWebGraphThermHygroAlert3 TRAP-TYPE ENTERPRISE wtWebGraphThermHygro VARIABLES { wtWebGraphThermHygroAlarmTrapText } DESCRIPTION "Dieser SNMP-Trap wird generiert, wenn der Alarm 3 ausgel�st wurde, das SNMP-Trap-Enable gesetzt ist und SNMP in Enable freigegeben ist." -- NMS trap annotation --#TYPE "Thermo-Hygrometer Alarm 3" --#SUMMARY "Alarm 3: %s" --#ARGUMENTS {0} --#SEVERITY INFORMATIONAL --#TIMEINDEX 99 --#STATE OPERATIONAL ::= 33 wtWebGraphThermHygroAlert4 TRAP-TYPE ENTERPRISE wtWebGraphThermHygro VARIABLES { wtWebGraphThermHygroAlarmTrapText } DESCRIPTION "Dieser SNMP-Trap wird generiert, wenn der Alarm 4 ausgel�st wurde, das SNMP-Trap-Enable gesetzt ist und SNMP in Enable freigegeben ist." -- NMS trap annotation --#TYPE "Thermo-Hygrometer Alarm 4" --#SUMMARY "Alarm 4: %s" --#ARGUMENTS {0} --#SEVERITY INFORMATIONAL --#TIMEINDEX 99 --#STATE OPERATIONAL ::= 34 wtWebGraphThermHygroAlert5 TRAP-TYPE ENTERPRISE wtWebGraphThermHygro VARIABLES { wtWebGraphThermHygroAlarmTrapText } DESCRIPTION "Dieser SNMP-Trap wird generiert, wenn der Alarm 5 ausgel�st wurde, das SNMP-Trap-Enable gesetzt ist und SNMP in Enable freigegeben ist." -- NMS trap annotation --#TYPE "Thermo-Hygrometer Alarm 5" --#SUMMARY "Alarm 5: %s" --#ARGUMENTS {0} --#SEVERITY INFORMATIONAL --#TIMEINDEX 99 --#STATE OPERATIONAL ::= 35 wtWebGraphThermHygroAlert6 TRAP-TYPE ENTERPRISE wtWebGraphThermHygro VARIABLES { wtWebGraphThermHygroAlarmTrapText } DESCRIPTION "Dieser SNMP-Trap wird generiert, wenn der Alarm 6 ausgel�st wurde, das SNMP-Trap-Enable gesetzt ist und SNMP in Enable freigegeben ist." -- NMS trap annotation --#TYPE "Thermo-Hygrometer Alarm 6" --#SUMMARY "Alarm 6: %s" --#ARGUMENTS {0} --#SEVERITY INFORMATIONAL --#TIMEINDEX 99 --#STATE OPERATIONAL ::= 36 wtWebGraphThermHygroAlert7 TRAP-TYPE ENTERPRISE wtWebGraphThermHygro VARIABLES { wtWebGraphThermHygroAlarmTrapText } DESCRIPTION "Dieser SNMP-Trap wird generiert, wenn der Alarm 7 ausgel�st wurde, das SNMP-Trap-Enable gesetzt ist und SNMP in Enable freigegeben ist." -- NMS trap annotation --#TYPE "Thermo-Hygrometer Alarm 7" --#SUMMARY "Alarm 7: %s" --#ARGUMENTS {0} --#SEVERITY INFORMATIONAL --#TIMEINDEX 99 --#STATE OPERATIONAL ::= 37 wtWebGraphThermHygroAlert8 TRAP-TYPE ENTERPRISE wtWebGraphThermHygro VARIABLES { wtWebGraphThermHygroAlarmTrapText } DESCRIPTION "Dieser SNMP-Trap wird generiert, wenn der Alarm 8 ausgel�st wurde, das SNMP-Trap-Enable gesetzt ist und SNMP in Enable freigegeben ist." -- NMS trap annotation --#TYPE "Thermo-Hygrometer Alarm 8" --#SUMMARY "Alarm 8: %s" --#ARGUMENTS {0} --#SEVERITY INFORMATIONAL --#TIMEINDEX 99 --#STATE OPERATIONAL ::= 38 wtWebGraphThermHygroAlert9 TRAP-TYPE ENTERPRISE wtWebGraphThermHygro VARIABLES { wtWebGraphThermHygroAlarmClearTrapText } DESCRIPTION "Dieser SNMP-Trap wird generiert, wenn der Alarm 1 gel�scht wurde, das SNMP-Trap-Enable gesetzt ist und SNMP in Enable freigegeben ist." -- NMS trap annotation --#TYPE "Thermo-Hygrometer Release Alarm 1" --#SUMMARY "Release Alarm 1: %s" --#ARGUMENTS {0} --#SEVERITY INFORMATIONAL --#TIMEINDEX 99 --#STATE OPERATIONAL ::= 91 wtWebGraphThermHygroAlert10 TRAP-TYPE ENTERPRISE wtWebGraphThermHygro VARIABLES { wtWebGraphThermHygroAlarmClearTrapText } DESCRIPTION "Dieser SNMP-Trap wird generiert, wenn der Alarm 2 gel�scht wurde, das SNMP-Trap-Enable gesetzt ist und SNMP in Enable freigegeben ist." -- NMS trap annotation --#TYPE "Thermo-Hygrometer Release Alarm 2" --#SUMMARY "Release Alarm 2: %s" --#ARGUMENTS {0} --#SEVERITY INFORMATIONAL --#TIMEINDEX 99 --#STATE OPERATIONAL ::= 92 wtWebGraphThermHygroAlert11 TRAP-TYPE ENTERPRISE wtWebGraphThermHygro VARIABLES { wtWebGraphThermHygroAlarmClearTrapText } DESCRIPTION "Dieser SNMP-Trap wird generiert, wenn der Alarm 3 gel�scht wurde, das SNMP-Trap-Enable gesetzt ist und SNMP in Enable freigegeben ist." -- NMS trap annotation --#TYPE "Thermo-Hygrometer Release Alarm 3" --#SUMMARY "Release Alarm 3: %s" --#ARGUMENTS {0} --#SEVERITY INFORMATIONAL --#TIMEINDEX 99 --#STATE OPERATIONAL ::= 93 wtWebGraphThermHygroAlert12 TRAP-TYPE ENTERPRISE wtWebGraphThermHygro VARIABLES { wtWebGraphThermHygroAlarmClearTrapText } DESCRIPTION "Dieser SNMP-Trap wird generiert, wenn der Alarm 4 gel�scht wurde, das SNMP-Trap-Enable gesetzt ist und SNMP in Enable freigegeben ist." -- NMS trap annotation --#TYPE "Thermo-Hygrometer Release Alarm 4" --#SUMMARY "Release Alarm 4: %s" --#ARGUMENTS {0} --#SEVERITY INFORMATIONAL --#TIMEINDEX 99 --#STATE OPERATIONAL ::= 94 wtWebGraphThermHygroAlert13 TRAP-TYPE ENTERPRISE wtWebGraphThermHygro VARIABLES { wtWebGraphThermHygroAlarmClearTrapText } DESCRIPTION "Dieser SNMP-Trap wird generiert, wenn der Alarm 5 gel�scht wurde, das SNMP-Trap-Enable gesetzt ist und SNMP in Enable freigegeben ist." -- NMS trap annotation --#TYPE "Thermo-Hygrometer Release Alarm 5" --#SUMMARY "Release Alarm 5: %s" --#ARGUMENTS {0} --#SEVERITY INFORMATIONAL --#TIMEINDEX 99 --#STATE OPERATIONAL ::= 95 wtWebGraphThermHygroAlert14 TRAP-TYPE ENTERPRISE wtWebGraphThermHygro VARIABLES { wtWebGraphThermHygroAlarmClearTrapText } DESCRIPTION "Dieser SNMP-Trap wird generiert, wenn der Alarm 6 gel�scht wurde, das SNMP-Trap-Enable gesetzt ist und SNMP in Enable freigegeben ist." -- NMS trap annotation --#TYPE "Thermo-Hygrometer Release Alarm 6" --#SUMMARY "Release Alarm 6: %s" --#ARGUMENTS {0} --#SEVERITY INFORMATIONAL --#TIMEINDEX 99 --#STATE OPERATIONAL ::= 96 wtWebGraphThermHygroAlert15 TRAP-TYPE ENTERPRISE wtWebGraphThermHygro VARIABLES { wtWebGraphThermHygroAlarmClearTrapText } DESCRIPTION "Dieser SNMP-Trap wird generiert, wenn der Alarm 7 gel�scht wurde, das SNMP-Trap-Enable gesetzt ist und SNMP in Enable freigegeben ist." -- NMS trap annotation --#TYPE "Thermo-Hygrometer Release Alarm 7" --#SUMMARY "Release Alarm 7: %s" --#ARGUMENTS {0} --#SEVERITY INFORMATIONAL --#TIMEINDEX 99 --#STATE OPERATIONAL ::= 97 wtWebGraphThermHygroAlert16 TRAP-TYPE ENTERPRISE wtWebGraphThermHygro VARIABLES { wtWebGraphThermHygroAlarmClearTrapText } DESCRIPTION "Dieser SNMP-Trap wird generiert, wenn der Alarm 8 gel�scht wurde, das SNMP-Trap-Enable gesetzt ist und SNMP in Enable freigegeben ist." -- NMS trap annotation --#TYPE "Thermo-Hygrometer Release Alarm 8" --#SUMMARY "Release Alarm 8: %s" --#ARGUMENTS {0} --#SEVERITY INFORMATIONAL --#TIMEINDEX 99 --#STATE OPERATIONAL ::= 98 wtWebGraphThermHygroAlertDiag TRAP-TYPE ENTERPRISE wtWebGraphThermHygro VARIABLES { wtWebGraphThermHygroDiagErrorIndex, wtWebGraphThermHygroDiagErrorMessage } DESCRIPTION "Dieser SNMP-Trap wird generiert, wenn eine Diagnose/Error-Message ausgel�st wurde, das SNMP-Trap-Enable gesetzt ist und SNMP in Enable freigegeben ist." -- NMS trap annotation --#TYPE "Thermo-Hygrometer Diagnose Error" --#SUMMARY "Diagnose error index: %d, Message: %s." --#ARGUMENTS {0,1} --#SEVERITY MAJOR --#TIMEINDEX 99 --#STATE DEGRADED ::= 110 END