mibs/MIBS/wut/WebGraph-Thermo-Hygrometer-MIB
2023-12-05 12:25:34 +01:00

2216 lines
72 KiB
Plaintext
Raw Blame History

WebGraph-Thermo-Hygrometer-MIB DEFINITIONS ::= BEGIN
----------------------------------------------------------------
-- Revison: 1.26
----------------------------------------------------------------
-- 03.08.2006 wtWebGraphThermHygroSensorNo korrigiert
--
----------------------------------------------------------------
-- Revison: 1.25
----------------------------------------------------------------
-- 28.06.2006 New: Community String Trap
--
----------------------------------------------------------------
-- Revison: 1.24
----------------------------------------------------------------
-- 26.07.2005 #57606, WebGraph Thermo-Hygrometer
-- CRON Zeitfunktion
----------------------------------------------------------------
-- Revison: 1.23
----------------------------------------------------------------
-- 14.07.2005 #57606, WebGraph Thermo-Hygrometer
-- FTP Funktionen
----------------------------------------------------------------
-- Revison: 1.22
----------------------------------------------------------------
-- 17.05.2005 #57606, WebGraph Thermo-Hygrometer
-- Trap NMS annotations
----------------------------------------------------------------
-- Revison: 1.21
----------------------------------------------------------------
-- 19.04.2005 #57606, WebGraph Thermo-Hygrometer
-- Mail Authentication
----------------------------------------------------------------
-- Revison: 1.20
----------------------------------------------------------------
-- 05.11.2004 #57606, WebGraph Themo-Hygrometer
-- Neu: Syslog
--
----------------------------------------------------------------
-- Revison: 1.19
----------------------------------------------------------------
-- 08.09.2004 #57606 WebGraph Thermo-Hygrometer
-- Neu:
-- wtWebGraphThermHygroAlarmRHMin
-- wtWebGraphThermHygroAlarmRHMax
-- wtWebGraphThermHygroAlarmRHHysteresis
-- wtWebGraphThermHygroPortSensorSelect
-- wtWebGraphThermHygroAlarmDeltaTemp
-- wtWebGraphThermHygroAlarmAHMin
-- wtWebGraphThermHygroAlarmAHMax
----------------------------------------------------------------
-- Revison: 1.18
----------------------------------------------------------------
-- 21.07.2004 #57601/3/4, Webio Digital
-- Neu:
-- Diagnosis
----------------------------------------------------------------
-- Revison: 1.17
----------------------------------------------------------------
-- 28.06.2004 #57630/1/2, Webio Digital
-- Neu: Puls Mode
----------------------------------------------------------------
-- Revison: 1.16
----------------------------------------------------------------
-- 16.04.2004 neu: #57632, Webio Digital EA24x24
--
-- #57630/1, Webio Digital EA12x12
-- Neu:
-- wtWebioEA12x12InputCounterClear
----------------------------------------------------------------
-- Revison: 1.15
----------------------------------------------------------------
-- 27.02.2004 #57630/1, Webio Digital EA12x12
-- Neu:
-- L<>nge erweitert auf 32 Bit:
-- wtWebioEA12x12InputValue
-- wtWebioEA12x12OutputValue
-- wtWebioEA12x12LoadControlView
-- wtWebioEA12x12LCShutDownView
-- wtWebioEA12x12SetOutput
-- wtWebioEA12x12AlarmInputTrigger
-- wtWebioEA12x12AlarmOutputTrigger
-- wtWebioEA12x12PortOutputSafetyState
----------------------------------------------------------------
-- Revison: 1.14
----------------------------------------------------------------
-- 16.02.2004 #57630/1, Webio Digital EA12x12
-- Neu:
-- Texte: Location, Contact
-- SNMP: ColdStart, WarmStart, ComunityString Read/RW
-- Alarm: ColdStart / WarmStart trigger
--
-- #57601, 1xWebThermometer
-- #57603, 2xWebThermometer
-- #57604, 8xWebThermometer
-- Neu:
-- Alarm Hysterese
-- Alarm Clear Traps und Mails
----------------------------------------------------------------
-- Revison: 1.13
----------------------------------------------------------------
-- 05.02.2004 #57601, 1xWebThermometer
-- #57603, 2xWebThermometer
-- #57604, 8xWebThermometer
-- Neu:
-- Texte: Location, Contact
-- SNMP: ColdStart, WarmStart, ComunityString Read/RW
-- Alarm: ColdStart / WarmStart trigger
----------------------------------------------------------------
-- Revison: 1.12
----------------------------------------------------------------
-- 05.11.2003 #57601, 1xWebThermometer
-- #57603, 2xWebThermometer
-- #57604, 8xWebThermometer
-- Neu:
-- Mail Option: CSV attachment
-- Offsetkalibrierung
-- Alarm: Tcp Client mode
----------------------------------------------------------------
-- Revison: 1.11
----------------------------------------------------------------
-- 28.10.2003 #57630/1, Webio Digital EA12x12
-- Neue Feature:
-- Alarm : TCP Client mode
-- INPUTS: Pulsfilter
----------------------------------------------------------------
-- Revison: 1.10
----------------------------------------------------------------
-- 04.07.2003 #57630, Webio Digital EA12x12
-- Neuer Men<65>baum, neue Bin<69>rmodi
--
----------------------------------------------------------------
-- Revison: 1.09
----------------------------------------------------------------
-- 24.03.2003 Trap Enterprise (Type 1, 2) korrigiert
--
----------------------------------------------------------------
-- Revison: 1.08
----------------------------------------------------------------
-- 04.03.2003 Abgleich der Sprachversionen DE/US
--
----------------------------------------------------------------
-- Revison: 1.07
----------------------------------------------------------------
-- 25.02.2003 Korrektur US: wtWebioEA12x12LoadControl
----------------------------------------------------------------
-- Revison: 1.06
----------------------------------------------------------------
-- 24.02.2003 Neu:
-- #57630, Webio EA12x12, LoadControl Group
-- Korrektur: Trap variable Namen
----------------------------------------------------------------
-- Revison: 1.05
----------------------------------------------------------------
-- 21.01.2003 Neu:
-- #57630, Webio EA12x12
----------------------------------------------------------------
-- Revison: 1.04
----------------------------------------------------------------
-- 02.01.2003 Neu:
-- #57601, 1xWebThermometer
-- #57603, 2xWebThermometer, 8 Alarme
----------------------------------------------------------------
-- Revison: 1.03
----------------------------------------------------------------
-- 21.11.2002 Neu:
-- #57604, 2xWebThermometer, BOOTP, StartupSite
-- #57603, 8xWebThermometer, BOOTP, StartupSite
----------------------------------------------------------------
-- Revison: 1.02
----------------------------------------------------------------
-- 26.07.2002 Neu:
-- #57604, 2xWebThermometer
-- #57603, 8xWebThermometer
--
IMPORTS
OBJECT-TYPE FROM RFC-1212
TRAP-TYPE FROM RFC-1215
DisplayString, PhysAddress FROM RFC1213-MIB
enterprises, TimeTicks, IpAddress, NetworkAddress FROM RFC1155-SMI;
-- IpAddress ::=
-- [APPLICATION 0]
-- IMPLICIT OCTET STRING (SIZE (4))
-- Counter ::=
-- [APPLICATION 1]
-- IMPLICIT INTEGER (0..4294967295)
-- Gauge ::=
-- [APPLICATION 2]
-- IMPLICIT INTEGER (0..4294967295)
-- TimeTicks ::=
-- [APPLICATION 3]
-- IMPLICIT INTEGER (0..4294967295)
-- DisplayString ::= OCTET STRING
-- PhysAddress ::= OCTET STRING
-- internet OBJECT IDENTIFIER ::= { iso org(3) dod(6) 1 }
-- privat OBJECT IDENTIFIER ::= { internet 4 }
-- enterprises OBJECT IDENTIFIER ::= { privat 1 }
-- Wiesemann & Theis Com-Server MIB
wut OBJECT IDENTIFIER ::= { enterprises 5040 }
wtComServer OBJECT IDENTIFIER ::= { wut 1 }
wtWebio OBJECT IDENTIFIER ::= { wtComServer 2 }
-- wtWebGraph Tree
wtWebGraphThermHygro OBJECT IDENTIFIER ::= { wtWebio 9 }
-- wtWebGraphThermHygro Tree
wtWebGraphThermHygroTemp OBJECT IDENTIFIER ::= { wtWebGraphThermHygro 1 }
wtWebGraphThermHygroSessCntrl OBJECT IDENTIFIER ::= { wtWebGraphThermHygro 2 }
wtWebGraphThermHygroConfig OBJECT IDENTIFIER ::= { wtWebGraphThermHygro 3 }
wtWebGraphThermHygroDiag OBJECT IDENTIFIER ::= { wtWebGraphThermHygro 4 }
-- wtWebGraphThermHygroConfig Tree
wtWebGraphThermHygroDevice OBJECT IDENTIFIER ::= { wtWebGraphThermHygroConfig 1 }
wtWebGraphThermHygroPorts OBJECT IDENTIFIER ::= { wtWebGraphThermHygroConfig 2 }
wtWebGraphThermHygroManufact OBJECT IDENTIFIER ::= { wtWebGraphThermHygroConfig 3 }
-- wtWebGraphThermHygroDevice Tree
wtWebGraphThermHygroText OBJECT IDENTIFIER ::= { wtWebGraphThermHygroDevice 1 }
wtWebGraphThermHygroTimeDate OBJECT IDENTIFIER ::= { wtWebGraphThermHygroDevice 2 }
wtWebGraphThermHygroBasic OBJECT IDENTIFIER ::= { wtWebGraphThermHygroDevice 3 }
wtWebGraphThermHygroDatalogger OBJECT IDENTIFIER ::= { wtWebGraphThermHygroDevice 4 }
wtWebGraphThermHygroAlarm OBJECT IDENTIFIER ::= { wtWebGraphThermHygroDevice 5 }
-- wtWebGraphThermHygroTimeDate Tree
wtWebGraphThermHygroTimeZone OBJECT IDENTIFIER ::= { wtWebGraphThermHygroTimeDate 1 }
wtWebGraphThermHygroTimeServer OBJECT IDENTIFIER ::= { wtWebGraphThermHygroTimeDate 2 }
wtWebGraphThermHygroDeviceClock OBJECT IDENTIFIER ::= { wtWebGraphThermHygroTimeDate 3 }
-- wtWebGraphThermHygroBasic Tree
wtWebGraphThermHygroNetwork OBJECT IDENTIFIER ::= { wtWebGraphThermHygroBasic 1 }
wtWebGraphThermHygroHTTP OBJECT IDENTIFIER ::= { wtWebGraphThermHygroBasic 2 }
wtWebGraphThermHygroMail OBJECT IDENTIFIER ::= { wtWebGraphThermHygroBasic 3 }
wtWebGraphThermHygroSNMP OBJECT IDENTIFIER ::= { wtWebGraphThermHygroBasic 4 }
wtWebGraphThermHygroUDP OBJECT IDENTIFIER ::= { wtWebGraphThermHygroBasic 5 }
wtWebGraphThermHygroSyslog OBJECT IDENTIFIER ::= { wtWebGraphThermHygroBasic 6 }
wtWebGraphThermHygroFTP OBJECT IDENTIFIER ::= { wtWebGraphThermHygroBasic 7 }
-- wtWebGraphThermHygroTemp Group
------------------------
wtWebGraphThermHygroSensors OBJECT-TYPE
SYNTAX INTEGER (1..2)
ACCESS read-only
STATUS mandatory
DESCRIPTION
"Anzahl der Sensoren"
::= { wtWebGraphThermHygroTemp 1 }
-- Interface Table
wtWebGraphThermHygroSensorTable OBJECT-TYPE
SYNTAX SEQUENCE OF WtWebGraphThermHygroSensorEntry
ACCESS not-accessible
STATUS mandatory
DESCRIPTION
"Liste der Sensoren."
::= { wtWebGraphThermHygroTemp 2 }
wtWebGraphThermHygroSensorEntry OBJECT-TYPE
SYNTAX WtWebGraphThermHygroSensorEntry
ACCESS not-accessible
STATUS mandatory
DESCRIPTION
"Nummern der Sensoren."
INDEX { wtWebGraphThermHygroSensorNo }
::= { wtWebGraphThermHygroSensorTable 1 }
WtWebGraphThermHygroSensorEntry ::= SEQUENCE {
wtWebGraphThermHygroSensorNo INTEGER
}
wtWebGraphThermHygroSensorNo OBJECT-TYPE
SYNTAX INTEGER (1..2)
ACCESS read-only
STATUS mandatory
DESCRIPTION
"Nummer des Sensors."
::= { wtWebGraphThermHygroSensorEntry 1 }
-- wtWebGraphThermHygroTempValue Table
wtWebGraphThermHygroTempValueTable OBJECT-TYPE
SYNTAX SEQUENCE OF WtWebGraphThermHygroTempValueEntry
ACCESS not-accessible
STATUS mandatory
DESCRIPTION
"Liste Sensormesswerte korrespondiernd zu den Kanalnummern."
::= { wtWebGraphThermHygroTemp 3 }
wtWebGraphThermHygroTempValueEntry OBJECT-TYPE
SYNTAX WtWebGraphThermHygroTempValueEntry
ACCESS not-accessible
STATUS mandatory
DESCRIPTION
"Messwerte der Sensoren."
INDEX { wtWebGraphThermHygroSensorNo }
::= { wtWebGraphThermHygroTempValueTable 1 }
WtWebGraphThermHygroTempValueEntry ::= SEQUENCE{
wtWebGraphThermHygroTempValue OCTET STRING
}
wtWebGraphThermHygroTempValue OBJECT-TYPE
SYNTAX OCTET STRING (SIZE (5))
ACCESS read-only
STATUS mandatory
DESCRIPTION
"Messwert des Sensors, ASCII, 5 Zeichen lang."
::= { wtWebGraphThermHygroTempValueEntry 1 }
-- wtWebGraphThermHygroBinaryTempValue Table
wtWebGraphThermHygroBinaryTempValueTable OBJECT-TYPE
SYNTAX SEQUENCE OF WtWebGraphThermHygroBinaryTempValueEntry
ACCESS not-accessible
STATUS mandatory
DESCRIPTION
"Liste bin<69>rer Messwerte korrespondiernd zu den Kanalnummern."
::= { wtWebGraphThermHygroTemp 4 }
wtWebGraphThermHygroBinaryTempValueEntry OBJECT-TYPE
SYNTAX WtWebGraphThermHygroBinaryTempValueEntry
ACCESS not-accessible
STATUS mandatory
DESCRIPTION
"Bin<69>re Messwerte der Sensoren."
INDEX { wtWebGraphThermHygroSensorNo }
::= { wtWebGraphThermHygroBinaryTempValueTable 1 }
WtWebGraphThermHygroBinaryTempValueEntry ::= SEQUENCE{
wtWebGraphThermHygroBinaryTempValue INTEGER
}
wtWebGraphThermHygroBinaryTempValue OBJECT-TYPE
SYNTAX INTEGER
ACCESS read-only
STATUS mandatory
DESCRIPTION
"Bin<69>rer Messwert des Sensors in 1/10 Grad Celcius oder 1/10 Prozent,
vorzeichenbehaftete Zahl, der Wert 0x4FFF6 zeigt einen
ung<6E>ltigen Messwert an."
::= { wtWebGraphThermHygroBinaryTempValueEntry 1 }
-- wtWebGraphThermHygroSessCntrl Group
--------------------------------
wtWebGraphThermHygroSessCntrlPassword OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Durch Eintrag diese Passworts wird eine Session er<65>ffnet, die durch Schreiben einer 1
in wtWebGraphThermHygroSessCntrlLogout beendet wird. Gelesen liefert diese Variable einen String der
L<>nge Null."
::= { wtWebGraphThermHygroSessCntrl 1 }
wtWebGraphThermHygroSessCntrlConfigMode OBJECT-TYPE
SYNTAX INTEGER {
wtWebGraphThermHygroSessCntrl-NoSession(0),
wtWebGraphThermHygroSessCntrl-Session(1)
}
ACCESS read-only
STATUS mandatory
DESCRIPTION
"Wenn sich das Ger<65>t im SNMP-Konfigurationsmodus befindet, wird der Wert
1 zur<75>ckgeliefert, ansonsten 0. Ist der ausgelesene Wert 0
obwohl das Passwort geschrieben wurde, befindet sich das Ger<65>t bereits
im Configmodus und die Variablen k<>nnen nur gelesen werden."
::= { wtWebGraphThermHygroSessCntrl 2 }
wtWebGraphThermHygroSessCntrlLogout OBJECT-TYPE
SYNTAX INTEGER
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Wenn sich das Ger<65>t im Konfigurationsmodus befindet und die Session wurde
von diesem SNMP Manager er<65>ffnet, kann durch Schreiben einer 1 in diese
Variable die Session beendet werden und alle Parameter werden gespeichert.
Durch Schreiben einer 2 werden alle Eingaben wieder verworfen.
Gelesen liefert diese Variable den Wert 0."
::= { wtWebGraphThermHygroSessCntrl 3 }
wtWebGraphThermHygroSessCntrlAdminPassword OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Dieses Passwort sch<63>tzt den Zugang der Variablen mit Administratorrechten. Es kann nur vom
Administrator ver<65>ndert werden. Wurde kein Passwort vergeben so ist public das g<>ltige Passwort.
Gelesen liefert diese Variable einen Null-Pointer."
::= { wtWebGraphThermHygroSessCntrl 4 }
wtWebGraphThermHygroSessCntrlConfigPassword OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Dieses Passwort sch<63>tzt den Zugang der Variablen mit Configuratorrechten. Es kann vom
Administrator und vom Configurator ver<65>ndert werden. Wurde kein Passwort vergeben so ist
public das g<>ltige Passwort.Gelesen liefert diese Variable einen Null-Pointer."
::= { wtWebGraphThermHygroSessCntrl 5 }
-- wtWebGraphThermHygroText Group
------------------------
wtWebGraphThermHygroDeviceName OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Name des Webio's, der userdefinierbar ist."
::= { wtWebGraphThermHygroText 1 }
wtWebGraphThermHygroDeviceText OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Beschreibung des Webio's, die userdefinierbar ist."
::= { wtWebGraphThermHygroText 2 }
wtWebGraphThermHygroDeviceLocation OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Aufstellort des Webio's, die userdefinierbar ist."
::= { wtWebGraphThermHygroText 3 }
wtWebGraphThermHygroDeviceContact OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Kontaktadresse im Servicefall des Webio's, die userdefinierbar ist."
::= { wtWebGraphThermHygroText 4 }
-- wtWebGraphThermHygroTimeZone Group
----------------------------
wtWebGraphThermHygroTzOffsetHrs OBJECT-TYPE
SYNTAX INTEGER
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Offset Stunden der Zeitzone (Winterzeit) in Bezug auf UTC."
::= { wtWebGraphThermHygroTimeZone 1 }
wtWebGraphThermHygroTzOffsetMin OBJECT-TYPE
SYNTAX INTEGER
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Offset Minuten der Zeitzone (Winterzeit) in Bezug auf UTC."
::= { wtWebGraphThermHygroTimeZone 2 }
wtWebGraphThermHygroTzEnable OBJECT-TYPE
SYNTAX OCTET STRING (SIZE(4))
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Bin<69>r kodierte Bedingung:
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3: ungenutzt
Octet 4:
Bit 1-7: ungenutzt
Bit 0 : Apply Timezone"
::= { wtWebGraphThermHygroTimeZone 3 }
wtWebGraphThermHygroStTzOffsetHrs OBJECT-TYPE
SYNTAX INTEGER
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Offset Stunden der Zeitzone (Sommerzeit) in Bezug auf UTC."
::= { wtWebGraphThermHygroTimeZone 4 }
wtWebGraphThermHygroStTzOffsetMin OBJECT-TYPE
SYNTAX INTEGER
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Offset Minuten der Zeitzone (Sommerzeit) in Bezug auf UTC."
::= { wtWebGraphThermHygroTimeZone 5 }
wtWebGraphThermHygroStTzEnable OBJECT-TYPE
SYNTAX OCTET STRING (SIZE(4))
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Bin<69>r kodierte Bedingung:
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3: ungenutzt
Octet 4:
Bit 1-7: ungenutzt
Bit 0 : Apply Summertime"
::= { wtWebGraphThermHygroTimeZone 6 }
wtWebGraphThermHygroStTzStartMonth OBJECT-TYPE
SYNTAX INTEGER {
wtWebGraphThermHygroStartMonth-January(1),
wtWebGraphThermHygroStartMonth-February(2),
wtWebGraphThermHygroStartMonth-March(3),
wtWebGraphThermHygroStartMonth-April(4),
wtWebGraphThermHygroStartMonth-May(5),
wtWebGraphThermHygroStartMonth-June(6),
wtWebGraphThermHygroStartMonth-July(7),
wtWebGraphThermHygroStartMonth-August(8),
wtWebGraphThermHygroStartMonth-September(9),
wtWebGraphThermHygroStartMonth-October(10),
wtWebGraphThermHygroStartMonth-November(11),
wtWebGraphThermHygroStartMonth-December(12)
}
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Monat, in dem von Winterzeit auf Sommerzeit umgestellt wird."
::= { wtWebGraphThermHygroTimeZone 7 }
wtWebGraphThermHygroStTzStartMode OBJECT-TYPE
SYNTAX INTEGER {
wtWebGraphThermHygroStartMode-first(1),
wtWebGraphThermHygroStartMode-second(2),
wtWebGraphThermHygroStartMode-third(3),
wtWebGraphThermHygroStartMode-fourth(4),
wtWebGraphThermHygroStartMode-last(5)
}
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Modus des Wochentages, bei dem von Winterzeit auf Sommerzeit umgestellt wird."
::= { wtWebGraphThermHygroTimeZone 8 }
wtWebGraphThermHygroStTzStartWday OBJECT-TYPE
SYNTAX INTEGER {
wtWebGraphThermHygroStartWday-Sunday(1),
wtWebGraphThermHygroStartWday-Monday(2),
wtWebGraphThermHygroStartWday-Tuesday(3),
wtWebGraphThermHygroStartWday-Thursday(4),
wtWebGraphThermHygroStartWday-Wednesday(5),
wtWebGraphThermHygroStartWday-Friday(6),
wtWebGraphThermHygroStartWday-Saturday(7)
}
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Wochentag, bei dem von Winterzeit auf Sommerzeit umgestellt wird"
::= { wtWebGraphThermHygroTimeZone 9 }
wtWebGraphThermHygroStTzStartHrs OBJECT-TYPE
SYNTAX INTEGER
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Stunde, in der die Zeitzone von Winter auf Sommerzeit in Bezug auf UTC
umgestellt wird."
::= { wtWebGraphThermHygroTimeZone 10 }
wtWebGraphThermHygroStTzStartMin OBJECT-TYPE
SYNTAX INTEGER
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Minute, in der die Zeitzone von Winter auf Sommerzeit in Bezug auf UTC
umgestellt wird."
::= { wtWebGraphThermHygroTimeZone 11 }
wtWebGraphThermHygroStTzStopMonth OBJECT-TYPE
SYNTAX INTEGER {
wtWebGraphThermHygroStopMonth-January(1),
wtWebGraphThermHygroStopMonth-February(2),
wtWebGraphThermHygroStopMonth-March(3),
wtWebGraphThermHygroStopMonth-April(4),
wtWebGraphThermHygroStopMonth-May(5),
wtWebGraphThermHygroStopMonth-June(6),
wtWebGraphThermHygroStopMonth-July(7),
wtWebGraphThermHygroStopMonth-August(8),
wtWebGraphThermHygroStopMonth-September(9),
wtWebGraphThermHygroStopMonth-October(10),
wtWebGraphThermHygroStopMonth-November(11),
wtWebGraphThermHygroStopMonth-December(12)
}
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Monat, bei dem von Sommerzeit auf Winterzeit umgestellt wird."
::= { wtWebGraphThermHygroTimeZone 12 }
wtWebGraphThermHygroStTzStopMode OBJECT-TYPE
SYNTAX INTEGER {
wtWebGraphThermHygroStopMode-first(1),
wtWebGraphThermHygroStopMode-second(2),
wtWebGraphThermHygroStopMode-third(3),
wtWebGraphThermHygroStopMode-fourth(4),
wtWebGraphThermHygroStopMode-last(5)
}
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Modus des Wochentages, bei dem von Sommerzeit auf Winterzeit umgestellt wird."
::= { wtWebGraphThermHygroTimeZone 13 }
wtWebGraphThermHygroStTzStopWday OBJECT-TYPE
SYNTAX INTEGER {
wtWebGraphThermHygroStopWday-Sunday(1),
wtWebGraphThermHygroStopWday-Monday(2),
wtWebGraphThermHygroStopWday-Tuesday(3),
wtWebGraphThermHygroStopWday-Thursday(4),
wtWebGraphThermHygroStopWday-Wednesday(5),
wtWebGraphThermHygroStopWday-Friday(6),
wtWebGraphThermHygroStopWday-Saturday(7)
}
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Wochentage, bei dem von Sommerzeit auf Winterzeit umgestellt wird."
::= { wtWebGraphThermHygroTimeZone 14 }
wtWebGraphThermHygroStTzStopHrs OBJECT-TYPE
SYNTAX INTEGER
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Stunde, in der die Zeitzone von Sommer auf Winterzeit in Bezug auf UTC
umgestellt wird."
::= { wtWebGraphThermHygroTimeZone 15 }
wtWebGraphThermHygroStTzStopMin OBJECT-TYPE
SYNTAX INTEGER
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Minute, in der die Zeitzone von Sommer auf Winterzeit in Bezug auf UTC
umgestellt wird."
::= { wtWebGraphThermHygroTimeZone 16 }
-- wtWebGraphThermHygroTimeServer Group
----------------------------
wtWebGraphThermHygroTimeServer1 OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"IP-Adresse oder Name des 1. Time Servers in ASCII"
::= { wtWebGraphThermHygroTimeServer 1 }
wtWebGraphThermHygroTimeServer2 OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"IP-Adresse oder Name des 2. Time Servers in ASCII"
::= { wtWebGraphThermHygroTimeServer 2 }
wtWebGraphThermHygroTsEnable OBJECT-TYPE
SYNTAX OCTET STRING (SIZE(4))
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Bin<69>r kodierte Bedingung:
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3: ungenutzt
Octet 4:
Bit 1-7: ungenutzt
Bit 0 : Apply Time Server"
::= { wtWebGraphThermHygroTimeServer 3 }
wtWebGraphThermHygroTsSyncTime OBJECT-TYPE
SYNTAX INTEGER
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Die Abfrage der Timeserver erfolgt zu dieser vollen Stunde (0-23)."
::= { wtWebGraphThermHygroTimeServer 4 }
-- wtWebGraphThermHygroDeviceClock Group
----------------------------
wtWebGraphThermHygroClockHrs OBJECT-TYPE
SYNTAX INTEGER (0..23)
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Uhrzeit Stunde, unter Ber<65>cksichtigung der Zeitzoneneinstellungen."
::= { wtWebGraphThermHygroDeviceClock 1 }
wtWebGraphThermHygroClockMin OBJECT-TYPE
SYNTAX INTEGER (0..59)
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Uhrzeit Minute, unter Ber<65>cksichtigung der Zeitzoneneinstellungen."
::= { wtWebGraphThermHygroDeviceClock 2 }
wtWebGraphThermHygroClockDay OBJECT-TYPE
SYNTAX INTEGER (1..31)
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Datum: Tag des Monats, unter Ber<65>cksichtigung der Zeitzoneneinstellungen."
::= { wtWebGraphThermHygroDeviceClock 3 }
wtWebGraphThermHygroClockMonth OBJECT-TYPE
SYNTAX INTEGER {
wtWebGraphThermHygroClockMonth-January(1),
wtWebGraphThermHygroClockMonth-February(2),
wtWebGraphThermHygroClockMonth-March(3),
wtWebGraphThermHygroClockMonth-April(4),
wtWebGraphThermHygroClockMonth-May(5),
wtWebGraphThermHygroClockMonth-June(6),
wtWebGraphThermHygroClockMonth-July(7),
wtWebGraphThermHygroClockMonth-August(8),
wtWebGraphThermHygroClockMonth-September(9),
wtWebGraphThermHygroClockMonth-October(10),
wtWebGraphThermHygroClockMonth-November(11),
wtWebGraphThermHygroClockMonth-December(12)
}
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Datum: Monat, unter Ber<65>cksichtigung der Zeitzoneneinstellungen."
::= { wtWebGraphThermHygroDeviceClock 4 }
wtWebGraphThermHygroClockYear OBJECT-TYPE
SYNTAX INTEGER (0..65535)
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Datum: Jahr, unter Ber<65>cksichtigung der Zeitzoneneinstellungen."
::= { wtWebGraphThermHygroDeviceClock 5 }
-- wtWebGraphThermHygroNetwork Group
------------------------
wtWebGraphThermHygroIpAddress OBJECT-TYPE
SYNTAX IpAddress
ACCESS read-write
STATUS mandatory
DESCRIPTION
"IP-Adresse des Webio's"
::= { wtWebGraphThermHygroNetwork 1 }
wtWebGraphThermHygroSubnetMask OBJECT-TYPE
SYNTAX IpAddress
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Subnetmask des Webio's"
::= { wtWebGraphThermHygroNetwork 2 }
wtWebGraphThermHygroGateway OBJECT-TYPE
SYNTAX IpAddress
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Gateway des Webio's"
::= { wtWebGraphThermHygroNetwork 3 }
wtWebGraphThermHygroDnsServer1 OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"IP-Adresse des 1. DNS Servers in ASCII"
::= { wtWebGraphThermHygroNetwork 4 }
wtWebGraphThermHygroDnsServer2 OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"IP-Adresse des 2. DNS Servers in ASCII"
::= { wtWebGraphThermHygroNetwork 5 }
wtWebGraphThermHygroAddConfig OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Zusatzkonfiguration (1=EIN):
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3: ungenutzt
Octet 4:
Bit 0: BOOTP disable
"
::= { wtWebGraphThermHygroNetwork 6 }
-- wtWebGraphThermHygroHTTP Group
--------------------------------
wtWebGraphThermHygroStartup OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Legt die Startseite fest, nur eine Auswahl zul<75>ssig:
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3: ungenutzt
Octet 4:
Bit 2: user.htm
Bit 1: home.htm
Bit 0: index.htm"
::= { wtWebGraphThermHygroHTTP 1 }
wtWebGraphThermHygroGetHeaderEnable OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Wenn freigeschaltet, wird eine Erkennungssequenz nach
jeder GET-Anfrage der Antwort vorangestellt:
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3: ungenutzt
Octet 4:
Bit 1-7: ungenutzt
Bit 0 : GET Header enable"
::= { wtWebGraphThermHygroHTTP 2 }
wtWebGraphThermHygroHttpPort OBJECT-TYPE
SYNTAX INTEGER (1..65535)
ACCESS read-write
STATUS mandatory
DESCRIPTION
"HTTP Port, an den die TCP-Anfragen gesendet werden k<>nnen (Default: 80)."
::= { wtWebGraphThermHygroHTTP 3 }
-- wtWebGraphThermHygroMail Group
--------------------------------
wtWebGraphThermHygroMailAdName OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Ger<65>tename f<>r die Mail-Administration."
::= { wtWebGraphThermHygroMail 1 }
wtWebGraphThermHygroMailReply OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Reply-Adresse f<>r den Mailserver."
::= { wtWebGraphThermHygroMail 2 }
wtWebGraphThermHygroMailServer OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Name oder Ip-Adresse des Mailservers."
::= { wtWebGraphThermHygroMail 3 }
wtWebGraphThermHygroMailEnable OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Erlaubt das versenden von Mails:
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3: ungenutzt
Octet 4:
Bit 1-7: ungenutzt
Bit 0 : Mail enable"
::= { wtWebGraphThermHygroMail 4 }
wtWebGraphThermHygroMailAuthentication OBJECT-TYPE
SYNTAX OCTET STRING (SIZE (4))
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Auswahl des Authenticationmodi:
die OCTETS werden bitweise interpretiert, es darf nur ein Bit gesetzt sein.
Bit 0 entspricht dem LSB.
OCTET 1: ungenutzt
OCTET 2: ungenutzt
OCTET 3: ungenutzt
OCTET 4: Bit 3-7: ungenutzt
Bit 2 : POP3 vor SMTP
Bit 1 : ESMTP, Authentifizierung
Bit 0 : SMTP, keine Authentifizierung"
::= { wtWebGraphThermHygroMail 5 }
wtWebGraphThermHygroMailAuthUser OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Name des Postfaches / Postfachusers."
::= { wtWebGraphThermHygroMail 6 }
wtWebGraphThermHygroMailAuthPassword OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Passwort zur Identifizierung des Users."
::= { wtWebGraphThermHygroMail 7 }
wtWebGraphThermHygroMailPop3Server OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Name oder Ip-Adresse des POP3 Servers."
::= { wtWebGraphThermHygroMail 8 }
-- wtWebGraphThermHygroSNMP Group
--------------------------------
wtWebGraphThermHygroSnmpCommunityStringRead OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Community string read only"
::= { wtWebGraphThermHygroSNMP 1 }
wtWebGraphThermHygroSnmpCommunityStringReadWrite OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Community string read / write"
::= { wtWebGraphThermHygroSNMP 2 }
wtWebGraphThermHygroSystemTrapManagerIP OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Name oder IP-Adresse des SNMP-Managers, der den System Trap entgegen nimmt."
::= { wtWebGraphThermHygroSNMP 3 }
wtWebGraphThermHygroSystemTrapEnable OBJECT-TYPE
SYNTAX OCTET STRING (SIZE(4))
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Bin<69>r kodierte Triggerbedingungen:
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3: ungenutzt
Octet 4:
Bit 0: ColdStart
Bit 1: WarmStart
Bit 2-7: ungenutzt"
::= { wtWebGraphThermHygroSNMP 4 }
wtWebGraphThermHygroSnmpEnable OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Bit 0 schaltet SNMP frei:
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3: ungenutzt
Octet 4:
Bit 1-7: ungenutzt
Bit 0 : SNMP enable"
::= { wtWebGraphThermHygroSNMP 5 }
wtWebGraphThermHygroSnmpCommunityStringTrap OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Community string der Traps."
::= { wtWebGraphThermHygroSNMP 6 }
-- wtWebGraphThermHygroUDP Group
--------------------------------
wtWebGraphThermHygroUdpPort OBJECT-TYPE
SYNTAX INTEGER (1..65535)
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Udp Port, an den die Udp-Anfragen GET Single gesendet werden k<>nnen"
::= { wtWebGraphThermHygroUDP 1 }
wtWebGraphThermHygroUdpEnable OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Schaltet UDP frei:
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3: ungenutzt
Octet 4:
Bit 1-7: ungenutzt
Bit 0 : UDP enable"
::= { wtWebGraphThermHygroUDP 2 }
-- wtWebGraphThermHygroSyslog Group
--------------------------------
wtWebGraphThermHygroSyslogServerIP OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Name oder IP-Adresse des Syslog-Servers, der die Syslog-Nachricht entgegen nimmt."
::= { wtWebGraphThermHygroSyslog 1 }
wtWebGraphThermHygroSyslogServerPort OBJECT-TYPE
SYNTAX INTEGER
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Portnummer des Syslog-Servers, default ist 514."
::= { wtWebGraphThermHygroSyslog 2 }
wtWebGraphThermHygroSyslogSystemMessagesEnable OBJECT-TYPE
SYNTAX OCTET STRING (SIZE(4))
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Bin<69>r kodierte Triggerbedingungen:
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3: ungenutzt
Octet 4:
Bit 0: ColdStart
Bit 1: WarmStart
Bit 2: Diag Messages
Bit 3-7: ungenutzt"
::= { wtWebGraphThermHygroSyslog 3 }
wtWebGraphThermHygroSyslogEnable OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Bit 0 schaltet Syslog frei:
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3: ungenutzt
Octet 4:
Bit 1-7: ungenutzt
Bit 0 : Syslog enable"
::= { wtWebGraphThermHygroSyslog 4 }
-- wtWebGraphThermHygroFTP Group
--------------------------------
wtWebGraphThermHygroFTPServerIP OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Name oder IP-Adresse des FTP-Servers, der die FTP-Nachricht entgegen nimmt."
::= { wtWebGraphThermHygroFTP 1 }
wtWebGraphThermHygroFTPServerControlPort OBJECT-TYPE
SYNTAX INTEGER
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Portnummer des Kontrollports des FTP-Servers, default ist 21."
::= { wtWebGraphThermHygroFTP 2 }
wtWebGraphThermHygroFTPUserName OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"FTP Loggin: Name des Anwenders."
::= { wtWebGraphThermHygroFTP 3 }
wtWebGraphThermHygroFTPPassword OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"FTP Loggin: Passwort des Anwenders."
::= { wtWebGraphThermHygroFTP 4 }
wtWebGraphThermHygroFTPAccount OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"FTP Loggin: Account-Name des Anwenderkontos."
::= { wtWebGraphThermHygroFTP 5 }
wtWebGraphThermHygroFTPOption OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Bit 0 schaltet den FTP Server in den Passiv-Modus,
der Client kontrolliert die Datenverbindung allein:
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3: ungenutzt
Octet 4:
Bit 1-7: ungenutzt
Bit 0 : PASV enable"
::= { wtWebGraphThermHygroFTP 6 }
wtWebGraphThermHygroFTPEnable OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Bit 0 schaltet FTP frei:
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3: ungenutzt
Octet 4:
Bit 1-7: ungenutzt
Bit 0 : FTP enable"
::= { wtWebGraphThermHygroFTP 7 }
-- wtWebGraphThermHygroDatalogger Group
--------------------------------
wtWebGraphThermHygroLoggerTimebase OBJECT-TYPE
SYNTAX INTEGER {
wtWebGraphThermHygroDatalogger-1Min(1),
wtWebGraphThermHygroDatalogger-5Min(2),
wtWebGraphThermHygroDatalogger-15Min(3),
wtWebGraphThermHygroDatalogger-60Min(4)
}
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Zeitbasis des Dataloggers"
::= { wtWebGraphThermHygroDatalogger 1 }
wtWebGraphThermHygroLoggerSensorSel OBJECT-TYPE
SYNTAX OCTET STRING (SIZE(4))
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Bin<69>r kodierte Triggerbedingungen:
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3: ungenutzt
Octet 4:
Bit 0 : Sensor 1
Bit 1 : Sensor 2
Bit 2-7: ungenutzt"
::= { wtWebGraphThermHygroDatalogger 2 }
wtWebGraphThermHygroDisplaySensorSel OBJECT-TYPE
SYNTAX OCTET STRING (SIZE(4))
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Die hier ausgew<65>hlten Sensoren werden grafisch angezeigt:
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3: ungenutzt
Octet 4:
Bit 0: Sensor 1
Bit 1: Sensor 2
Bit 2-7: ungenutzt"
::= { wtWebGraphThermHygroDatalogger 3 }
-- Graph Sensor Color Table
wtWebGraphThermHygroSensorColorTable OBJECT-TYPE
SYNTAX SEQUENCE OF WtWebGraphThermHygroSensorColorEntry
ACCESS not-accessible
STATUS mandatory
DESCRIPTION
"Liste der Sensorfarben in der grafischen Darstellung."
::= { wtWebGraphThermHygroDatalogger 4 }
wtWebGraphThermHygroSensorColorEntry OBJECT-TYPE
SYNTAX WtWebGraphThermHygroSensorColorEntry
ACCESS not-accessible
STATUS mandatory
DESCRIPTION
"Nummern der Sensoren."
INDEX { wtWebGraphThermHygroSensorNo }
::= { wtWebGraphThermHygroSensorColorTable 1 }
WtWebGraphThermHygroSensorColorEntry ::= SEQUENCE {
wtWebGraphThermHygroSensorColor OCTET STRING
}
wtWebGraphThermHygroSensorColor OBJECT-TYPE
SYNTAX OCTET STRING (SIZE(3))
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Farbe des Sensors in der grafischen Darstellung."
::= { wtWebGraphThermHygroSensorColorEntry 1 }
wtWebGraphThermHygroAutoScaleEnable OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Bit 0 schaltet Vertical Auto Scale frei:
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3: ungenutzt
Octet 4:
Bit 1-7: ungenutzt
Bit 0 : Auto Scale enable"
::= { wtWebGraphThermHygroDatalogger 5 }
wtWebGraphThermHygroVerticalUpperLimit OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Definiert die maximal sichtbare Temperatur f<>r die grafische Darstellung.
Temperatur in Grad mit maximal 2 Nachkommastellen."
::= { wtWebGraphThermHygroDatalogger 6 }
wtWebGraphThermHygroVerticalLowerLimit OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Definiert die Minal sichtbare Temperatur f<>r die grafische Darstellung.
Temperatur in Grad mit maximal 2 Nachkommastellen."
::= { wtWebGraphThermHygroDatalogger 7 }
wtWebGraphThermHygroHorizontalZoom OBJECT-TYPE
SYNTAX INTEGER {
wtWebGraphThermHygroZoom-25Min(1),
wtWebGraphThermHygroZoom-75Min(2),
wtWebGraphThermHygroZoom-5Std(3),
wtWebGraphThermHygroZoom-30Std(4),
wtWebGraphThermHygroZoom-5Tage(5),
wtWebGraphThermHygroZoom-25Tage(6)
}
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Definiert den default Zoom."
::= { wtWebGraphThermHygroDatalogger 8 }
-- wtWebGraphThermHygroAlarm Group
----------------------------
wtWebGraphThermHygroAlarmCount OBJECT-TYPE
SYNTAX INTEGER (1..8)
ACCESS read-only
STATUS mandatory
DESCRIPTION
"Anzahl der unabh<62>ngigen Alarme."
::= { wtWebGraphThermHygroAlarm 1 }
-- Alarm Interface Table
wtWebGraphThermHygroAlarmIfTable OBJECT-TYPE
SYNTAX SEQUENCE OF WtWebGraphThermHygroAlarmIfEntry
ACCESS not-accessible
STATUS mandatory
DESCRIPTION
"Liste der Alarmindizes."
::= { wtWebGraphThermHygroAlarm 2 }
wtWebGraphThermHygroAlarmIfEntry OBJECT-TYPE
SYNTAX WtWebGraphThermHygroAlarmIfEntry
ACCESS not-accessible
STATUS mandatory
DESCRIPTION
"Nummern der Alarme."
INDEX { wtWebGraphThermHygroAlarmNo }
::= { wtWebGraphThermHygroAlarmIfTable 1 }
WtWebGraphThermHygroAlarmIfEntry ::= SEQUENCE {
wtWebGraphThermHygroAlarmNo INTEGER
}
wtWebGraphThermHygroAlarmNo OBJECT-TYPE
SYNTAX INTEGER (1..8)
ACCESS read-only
STATUS mandatory
DESCRIPTION
"Nummer des Alarms."
::= { wtWebGraphThermHygroAlarmIfEntry 1 }
-- Alarm Table
wtWebGraphThermHygroAlarmTable OBJECT-TYPE
SYNTAX SEQUENCE OF WtWebGraphThermHygroAlarmEntry
ACCESS not-accessible
STATUS mandatory
DESCRIPTION
"Liste der Einstellungen abh<62>ngig von der Alarmnummer."
::= { wtWebGraphThermHygroAlarm 3 }
wtWebGraphThermHygroAlarmEntry OBJECT-TYPE
SYNTAX WtWebGraphThermHygroAlarmEntry
ACCESS not-accessible
STATUS mandatory
DESCRIPTION
"Alarm-Parameter der einzelnen Alarme."
INDEX { wtWebGraphThermHygroAlarmNo }
::= { wtWebGraphThermHygroAlarmTable 1 }
WtWebGraphThermHygroAlarmEntry ::= SEQUENCE {
wtWebGraphThermHygroAlarmTrigger OCTET STRING,
wtWebGraphThermHygroAlarmMin OCTET STRING,
wtWebGraphThermHygroAlarmMax OCTET STRING,
wtWebGraphThermHygroAlarmHysteresis OCTET STRING,
wtWebGraphThermHygroAlarmDelay OCTET STRING,
wtWebGraphThermHygroAlarmInterval OCTET STRING,
wtWebGraphThermHygroAlarmEnable OCTET STRING,
wtWebGraphThermHygroAlarmEMailAddr OCTET STRING,
wtWebGraphThermHygroAlarmMailSubject OCTET STRING,
wtWebGraphThermHygroAlarmMailText OCTET STRING,
wtWebGraphThermHygroAlarmManagerIP OCTET STRING,
wtWebGraphThermHygroAlarmTrapText OCTET STRING,
wtWebGraphThermHygroAlarmMailOptions OCTET STRING,
wtWebGraphThermHygroAlarmTcpIpAddr OCTET STRING,
wtWebGraphThermHygroAlarmTcpPort INTEGER,
wtWebGraphThermHygroAlarmTcpText OCTET STRING,
wtWebGraphThermHygroAlarmClearMailSubject OCTET STRING,
wtWebGraphThermHygroAlarmClearMailText OCTET STRING,
wtWebGraphThermHygroAlarmClearTrapText OCTET STRING,
wtWebGraphThermHygroAlarmClearTcpText OCTET STRING,
wtWebGraphThermHygroAlarmDeltaTemp OCTET STRING,
wtWebGraphThermHygroAlarmRHMin OCTET STRING,
wtWebGraphThermHygroAlarmRHMax OCTET STRING,
wtWebGraphThermHygroAlarmRHHysteresis OCTET STRING,
wtWebGraphThermHygroAlarmAHMin OCTET STRING,
wtWebGraphThermHygroAlarmAHMax OCTET STRING,
wtWebGraphThermHygroAlarmSyslogIpAddr OCTET STRING,
wtWebGraphThermHygroAlarmSyslogPort INTEGER,
wtWebGraphThermHygroAlarmSyslogText OCTET STRING,
wtWebGraphThermHygroAlarmSyslogClearText OCTET STRING,
wtWebGraphThermHygroAlarmFtpDataPort OCTET STRING,
wtWebGraphThermHygroAlarmFtpFileName OCTET STRING,
wtWebGraphThermHygroAlarmFtpText OCTET STRING,
wtWebGraphThermHygroAlarmFtpClearText OCTET STRING,
wtWebGraphThermHygroAlarmFtpOption OCTET STRING,
wtWebGraphThermHygroAlarmTimerCron OCTET STRING
}
wtWebGraphThermHygroAlarmTrigger OBJECT-TYPE
SYNTAX OCTET STRING (SIZE(4))
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Bin<69>r kodierte Triggerbedingungen:
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3:
Bit 0: Timer
Bit 1: ColdStart
Bit 2: WarmStart
Bit 3: Sensor lost
Bit 4-7 ungenutzt
Octet 4:
Bit 0: Sensor 1
Bit 1: Sensor 2
Bit 2-7: ungenutzt"
::= { wtWebGraphThermHygroAlarmEntry 1 }
wtWebGraphThermHygroAlarmMin OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Minimum Temperatur in Grad mit maximal 2 Nachkommastellen."
::= { wtWebGraphThermHygroAlarmEntry 2 }
wtWebGraphThermHygroAlarmMax OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Maximum Temperatur in Grad mit maximal 2 Nachkommastellen."
::= { wtWebGraphThermHygroAlarmEntry 3 }
wtWebGraphThermHygroAlarmHysteresis OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Bei <20>berschreiten der Temperatur Min+Hyst. und Unterschreiten der Temperatur Max-Hyst.
wird der Alarm gel<65>scht: pos. Temperaturdifferenz in Grad mit maximal 2 Nachkommastellen."
::= { wtWebGraphThermHygroAlarmEntry 4 }
wtWebGraphThermHygroAlarmDelay OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Der Alarm wird erst ausgel<65>st, wenn f<>r die Dauer dieser Zeit
die Alarmbedingungen erf<72>llt sind (Zeit in Minuten)."
::= { wtWebGraphThermHygroAlarmEntry 5 }
wtWebGraphThermHygroAlarmInterval OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Sendeintervall von Alarmmeldungen."
::= { wtWebGraphThermHygroAlarmEntry 6 }
wtWebGraphThermHygroAlarmEnable OBJECT-TYPE
SYNTAX OCTET STRING (SIZE(4))
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Bin<69>r kodierte Alarmfreischaltung:
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3: ungenutzt
Octet 4:
Bit 0: Mail Enable
Bit 1: SNMP Trap enable
Bit 2: TCP Client enable"
::= { wtWebGraphThermHygroAlarmEntry 7 }
wtWebGraphThermHygroAlarmEMailAddr OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"E-Mail Adresse, an die die Alarmmail gesendet werden soll."
::= { wtWebGraphThermHygroAlarmEntry 8 }
wtWebGraphThermHygroAlarmMailSubject OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Betreffzeile der Alarmmail."
::= { wtWebGraphThermHygroAlarmEntry 9 }
wtWebGraphThermHygroAlarmMailText OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Text der Alarmmail: <Tn> liefert die aktuelle Temperatur; n = 1..8.
<Z> liefert die aktuelle Zeit."
::= { wtWebGraphThermHygroAlarmEntry 10 }
wtWebGraphThermHygroAlarmManagerIP OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Name oder IP-Adresse des SNMP-Managers, der den Trap entgegen nimmt."
::= { wtWebGraphThermHygroAlarmEntry 11 }
wtWebGraphThermHygroAlarmTrapText OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Dieser Text wird im Trapbody gesendet."
::= { wtWebGraphThermHygroAlarmEntry 12 }
wtWebGraphThermHygroAlarmMailOptions OBJECT-TYPE
SYNTAX OCTET STRING (SIZE(4))
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Bin<69>r kodierte Optionen:
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3: ungenutzt
Octet 4:
Bit 0 : Attach Thermo.csv Enable
Bit 1 : CSV-Data_since_last_report
Bit 2-7: ungenutzt"
::= { wtWebGraphThermHygroAlarmEntry 13 }
wtWebGraphThermHygroAlarmTcpIpAddr OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"IP-Adresse des Ger<65>tes, dass eine TCP-Nachricht empfangen soll."
::= { wtWebGraphThermHygroAlarmEntry 14 }
wtWebGraphThermHygroAlarmTcpPort OBJECT-TYPE
SYNTAX INTEGER (1..65535)
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Portnummer des Ger<65>tes, dass eine TCP-Nachricht empfangen soll."
::= { wtWebGraphThermHygroAlarmEntry 15 }
wtWebGraphThermHygroAlarmTcpText OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Dieser Text wird im TCP Trapbody gesendet."
::= { wtWebGraphThermHygroAlarmEntry 16 }
wtWebGraphThermHygroAlarmClearMailSubject OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Betreffzeile der Alarm Clear Mail."
::= { wtWebGraphThermHygroAlarmEntry 17 }
wtWebGraphThermHygroAlarmClearMailText OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Text der Alarm Clear Mail: <Tn> liefert die aktuelle Temperatur; n = 1..8.
<Z> liefert die aktuelle Zeit."
::= { wtWebGraphThermHygroAlarmEntry 18 }
wtWebGraphThermHygroAlarmClearTrapText OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Dieser Text wird im Trapbody gesendet."
::= { wtWebGraphThermHygroAlarmEntry 19 }
wtWebGraphThermHygroAlarmClearTcpText OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Dieser Text wird im TCP Trapbody gesendet."
::= { wtWebGraphThermHygroAlarmEntry 20 }
wtWebGraphThermHygroAlarmDeltaTemp OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Maximale Temperatur<75>nderung in Grad/min mit maximal 2 Nachkommastellen.
Es wird ein Mittelwert einer Periode von 5 Minuten verglichen."
::= { wtWebGraphThermHygroAlarmEntry 21 }
wtWebGraphThermHygroAlarmRHMin OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Minimum rel. Feuchte in Prozent mit maximal 2 Nachkommastellen."
::= { wtWebGraphThermHygroAlarmEntry 22 }
wtWebGraphThermHygroAlarmRHMax OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Maximum rel.Feuchte in Prozent mit maximal 2 Nachkommastellen."
::= { wtWebGraphThermHygroAlarmEntry 23 }
wtWebGraphThermHygroAlarmRHHysteresis OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Hysterese rel. Feuchte in Prozent mit maximal 2 Nachkommastellen."
::= { wtWebGraphThermHygroAlarmEntry 24 }
wtWebGraphThermHygroAlarmAHMin OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Minimum abs. Feuchte in g/(m*m*m) mit maximal 2 Nachkommastellen."
::= { wtWebGraphThermHygroAlarmEntry 25 }
wtWebGraphThermHygroAlarmAHMax OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Maximum abs. Feuchte in g/(m*m*m) mit maximal 2 Nachkommastellen."
::= { wtWebGraphThermHygroAlarmEntry 26 }
wtWebGraphThermHygroAlarmSyslogIpAddr OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"IP-Adresse des Ger<65>tes, dass eine Syslog-Nachricht empfangen soll."
::= { wtWebGraphThermHygroAlarmEntry 27 }
wtWebGraphThermHygroAlarmSyslogPort OBJECT-TYPE
SYNTAX INTEGER (1..65535)
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Portnummer des Ger<65>tes, dass eine Syslog-Nachricht empfangen soll, default Port ist 514."
::= { wtWebGraphThermHygroAlarmEntry 28 }
wtWebGraphThermHygroAlarmSyslogText OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Jede Syslog-Nachricht beginnt mit einem Priorit<69>tsschl<68>ssel, einem Zeitstempel und der Host Ip-Nummer gem<65><6D> RFC3164.
Dieser Text wird im darauf folgenden Syslog-Nachrichtenteil gesendet."
::= { wtWebGraphThermHygroAlarmEntry 29 }
wtWebGraphThermHygroAlarmSyslogClearText OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Jede Syslog-Nachricht beginnt mit einem Priorit<69>tsschl<68>ssel, einem Zeitstempel und der Host Ip-Nummer gem<65><6D> RFC3164.
Dieser Text wird im darauf folgenden Syslog-Nachrichtenteil gesendet."
::= { wtWebGraphThermHygroAlarmEntry 30 }
wtWebGraphThermHygroAlarmFtpDataPort OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Lokaler Datenport: 1..65535 oder AUTO: der n<>chste ungenutzte Port wird automatisch zugewiesen."
::= { wtWebGraphThermHygroAlarmEntry 31 }
wtWebGraphThermHygroAlarmFtpFileName OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Der File Name kann auch Pfadinformationen enthalten."
::= { wtWebGraphThermHygroAlarmEntry 32 }
wtWebGraphThermHygroAlarmFtpText OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Dieser Text wird in die angegebene Datei geschrieben."
::= { wtWebGraphThermHygroAlarmEntry 33 }
wtWebGraphThermHygroAlarmFtpClearText OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Dieser Text wird in die angegebene Datei geschrieben."
::= { wtWebGraphThermHygroAlarmEntry 34 }
wtWebGraphThermHygroAlarmFtpOptions OBJECT-TYPE
SYNTAX OCTET STRING (SIZE(4))
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Bin<69>r kodierte Optionen, nur eine Option darf eingeschaltet sein:
STORE = Datei ersetzen
APPEND = Nachrichten ans Dateiende anh<6E>ngen
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3: ungenutzt
Octet 4:
Bit 0 : STORE
Bit 1 : APPEND
Bit 2-7: ungenutzt"
::= { wtWebGraphThermHygroAlarmEntry 35 }
wtWebGraphThermHygroAlarmTimerCron OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Diser Eintrag enth<74>lt 5 Leerzeichen- oder Tab-getrente Felder zur Definition von zeitgesteuerten
Ausgabe von Alarmnachrichten. Jedes der 5 Felder muss mit der aktuelle Zeit <20>bereinstimmen, um
das Senden einer Nachricht auszul<75>sen, Wochentag und Monatstag werden getrennt ausgewertet.
Feld m<>glicher Zahlenbereich
1: Minuten 0-59
2. Stunden 0-23, 0 ist Mitternacht
3: Monatstag 1-31
4: Monat 1-12
5: Wochentag 0-6, 0 ist Sonntag
F<>r jedes Feld kann nun angegeben werden:
- eine ganze Zahl aus dem angegebenen Zahlenbereich
- ein Teil-Zahlenbereich (wie z.B. 1-5) innerhalb des angegebenen Zahlenbereichs
- eine mit Komma getrennte Liste von ganzen Zahlen oder Tei-Zahlenbereichen (wie z.B. 2,4-6,8,10)
- ein Stern * (deckt den gesamten erlaubten Bereich ab)
- ein / gibt die Schrittweite bei Teil-Zahlenbereichen an (z.B. 0-59/2 jede zweite Minute)."
::= { wtWebGraphThermHygroAlarmEntry 36 }
-- wtWebGraphThermHygroPorts Group
----------------------------
-- Table
wtWebGraphThermHygroPortTable OBJECT-TYPE
SYNTAX SEQUENCE OF WtWebGraphThermHygroPortEntry
ACCESS not-accessible
STATUS mandatory
DESCRIPTION
"Liste der Einstellungen abh<62>ngig von der Sensornummer."
::= { wtWebGraphThermHygroPorts 1 }
wtWebGraphThermHygroPortEntry OBJECT-TYPE
SYNTAX WtWebGraphThermHygroPortEntry
ACCESS not-accessible
STATUS mandatory
DESCRIPTION
"Sensor-Parameter der einzelnen Sensoren."
INDEX { wtWebGraphThermHygroSensorNo }
::= { wtWebGraphThermHygroPortTable 1 }
WtWebGraphThermHygroPortEntry ::= SEQUENCE {
wtWebGraphThermHygroPortName OCTET STRING,
wtWebGraphThermHygroPortText OCTET STRING,
wtWebGraphThermHygroPortOffset1 OCTET STRING,
wtWebGraphThermHygroPortTemperature1 OCTET STRING,
wtWebGraphThermHygroPortOffset2 OCTET STRING,
wtWebGraphThermHygroPortTemperature2 OCTET STRING,
wtWebGraphThermHygroPortComment OCTET STRING,
wtWebGraphThermHygroPortSensorSelect OCTET STRING
}
wtWebGraphThermHygroPortName OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Name des Sensors."
::= { wtWebGraphThermHygroPortEntry 1 }
wtWebGraphThermHygroPortText OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Beschreibung des Sensors."
::= { wtWebGraphThermHygroPortEntry 2 }
wtWebGraphThermHygroPortOffset1 OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Offset-Kalibrierung: Offset in Grad mit max. 2 Nachkommastellen bei Temperatur 1."
::= { wtWebGraphThermHygroPortEntry 3 }
wtWebGraphThermHygroPortTemperature1 OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Offset-Kalibrierung: Referenztemperatur 1 in Grad mit max. 2 Nachkommastellen."
::= { wtWebGraphThermHygroPortEntry 4 }
wtWebGraphThermHygroPortOffset2 OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Offset-Kalibrierung: Offset in Grad mit max. 2 Nachkommastellen bei Temperatur 2."
::= { wtWebGraphThermHygroPortEntry 5 }
wtWebGraphThermHygroPortTemperature2 OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Offset-Kalibrierung: Referenztemperatur 2 in Grad mit max. 2 Nachkommastellen."
::= { wtWebGraphThermHygroPortEntry 6 }
wtWebGraphThermHygroPortComment OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Offset-Kalibrierung: Angaben zu Datum, Referenzger<65>t, Operator."
::= { wtWebGraphThermHygroPortEntry 7 }
wtWebGraphThermHygroPortSensorSelect OBJECT-TYPE
SYNTAX OCTET STRING (SIZE(4))
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Bin<69>r kodierte Optionen f<>r Sensor 1:
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3: ungenutzt
Octet 4:
Bit 0 : PT100
Bit 1 : PT1000 (default)
Bit 2-7: ungenutzt
Bin<69>r kodierte Optionen f<>r Sensor 2:
Octet 1: ungenutzt
Octet 2: ungenutzt
Octet 3: ungenutzt
Octet 4:
Bit 0 : W&T Sensor rel. Feuchte (default)
Bit 1 : Skalar 0-2.5V
Bit 2 : Disconnect
Bit 3-7: ungenutzt"
::= { wtWebGraphThermHygroPortEntry 8 }
-- wtWebGraphThermHygroManufact Group
----------------------------
wtWebGraphThermHygroMfName OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Hersteller Firmenname"
::= { wtWebGraphThermHygroManufact 1 }
wtWebGraphThermHygroMfAddr OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Adresse des Herstellers."
::= { wtWebGraphThermHygroManufact 2 }
wtWebGraphThermHygroMfHotline OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Hotline des Herstellers."
::= { wtWebGraphThermHygroManufact 3 }
wtWebGraphThermHygroMfInternet OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Internet-Adresse des Herstellers."
::= { wtWebGraphThermHygroManufact 4 }
wtWebGraphThermHygroMfDeviceTyp OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Ger<65>tebezeichnung (Typ) des Herstellers."
::= { wtWebGraphThermHygroManufact 5 }
wtWebGraphThermHygroMfOrderNo OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Bestellnummer des Ger<65>tes des Herstellers."
::= { wtWebGraphThermHygroManufact 6 }
-- wtWebGraphThermHygroDiag Group
----------------------------
wtWebGraphThermHygroDiagErrorCount OBJECT-TYPE
SYNTAX INTEGER
ACCESS read-only
STATUS mandatory
DESCRIPTION
"Anzahl der Eintr<74>ge im Fehlerspeicher."
::= { wtWebGraphThermHygroDiag 1 }
wtWebGraphThermHygroDiagBinaryError OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-only
STATUS mandatory
DESCRIPTION
"Bitweise kodierte Fehlerursachen."
::= { wtWebGraphThermHygroDiag 2 }
wtWebGraphThermHygroDiagErrorIndex OBJECT-TYPE
SYNTAX INTEGER
ACCESS read-write
STATUS mandatory
DESCRIPTION
"Der Wert dieser Variable wird als Index f<>r die Ausgabe
eines Fehlertextes (wtWebGraphThermHygroDiagErrorMessage) verwendet.
Die Fehler im Speicher sind von 1 bis wtWebGraphThermHygroDiagErrorCount durchnummeriert.
Nach dem Schreiben einens g<>ltigen Wertes in diese Variable kann der dazugeh<65>rige Text
in wtWebGraphThermHygroDiagErrorMessage ausgelesen werden."
::= { wtWebGraphThermHygroDiag 3 }
wtWebGraphThermHygroDiagErrorMessage OBJECT-TYPE
SYNTAX OCTET STRING
ACCESS read-only
STATUS mandatory
DESCRIPTION
"Abh<62>ngig vom Wert wtWebGraphThermHygroDiagErrorIndex wird ein Fehlertext angezeigt."
::= { wtWebGraphThermHygroDiag 4 }
wtWebGraphThermHygroDiagErrorClear OBJECT-TYPE
SYNTAX INTEGER
ACCESS write-only
STATUS mandatory
DESCRIPTION
"Durch Schreiben eine 1 in diese Variable, wird der Fehlerspeicher gel<65>scht."
::= { wtWebGraphThermHygroDiag 5 }
-- TRAP DEFINITIONS
-- Diese SNMP-Traps werden gesendet, wenn die Alarmtriggerbedingungen zutreffen
wtWebGraphThermHygroAlert1 TRAP-TYPE
ENTERPRISE wtWebGraphThermHygro
VARIABLES { wtWebGraphThermHygroAlarmTrapText }
DESCRIPTION
"Dieser SNMP-Trap wird generiert, wenn der Alarm 1 ausgel<65>st wurde,
das SNMP-Trap-Enable gesetzt ist und
SNMP in Enable freigegeben ist."
-- NMS trap annotation
--#TYPE "Thermo-Hygrometer Alarm 1"
--#SUMMARY "Alarm 1: %s"
--#ARGUMENTS {0}
--#SEVERITY INFORMATIONAL
--#TIMEINDEX 99
--#STATE OPERATIONAL
::= 31
wtWebGraphThermHygroAlert2 TRAP-TYPE
ENTERPRISE wtWebGraphThermHygro
VARIABLES { wtWebGraphThermHygroAlarmTrapText }
DESCRIPTION
"Dieser SNMP-Trap wird generiert, wenn der Alarm 2 ausgel<65>st wurde,
das SNMP-Trap-Enable gesetzt ist und
SNMP in Enable freigegeben ist."
-- NMS trap annotation
--#TYPE "Thermo-Hygrometer Alarm 2"
--#SUMMARY "Alarm 2: %s"
--#ARGUMENTS {0}
--#SEVERITY INFORMATIONAL
--#TIMEINDEX 99
--#STATE OPERATIONAL
::= 32
wtWebGraphThermHygroAlert3 TRAP-TYPE
ENTERPRISE wtWebGraphThermHygro
VARIABLES { wtWebGraphThermHygroAlarmTrapText }
DESCRIPTION
"Dieser SNMP-Trap wird generiert, wenn der Alarm 3 ausgel<65>st wurde,
das SNMP-Trap-Enable gesetzt ist und
SNMP in Enable freigegeben ist."
-- NMS trap annotation
--#TYPE "Thermo-Hygrometer Alarm 3"
--#SUMMARY "Alarm 3: %s"
--#ARGUMENTS {0}
--#SEVERITY INFORMATIONAL
--#TIMEINDEX 99
--#STATE OPERATIONAL
::= 33
wtWebGraphThermHygroAlert4 TRAP-TYPE
ENTERPRISE wtWebGraphThermHygro
VARIABLES { wtWebGraphThermHygroAlarmTrapText }
DESCRIPTION
"Dieser SNMP-Trap wird generiert, wenn der Alarm 4 ausgel<65>st wurde,
das SNMP-Trap-Enable gesetzt ist und
SNMP in Enable freigegeben ist."
-- NMS trap annotation
--#TYPE "Thermo-Hygrometer Alarm 4"
--#SUMMARY "Alarm 4: %s"
--#ARGUMENTS {0}
--#SEVERITY INFORMATIONAL
--#TIMEINDEX 99
--#STATE OPERATIONAL
::= 34
wtWebGraphThermHygroAlert5 TRAP-TYPE
ENTERPRISE wtWebGraphThermHygro
VARIABLES { wtWebGraphThermHygroAlarmTrapText }
DESCRIPTION
"Dieser SNMP-Trap wird generiert, wenn der Alarm 5 ausgel<65>st wurde,
das SNMP-Trap-Enable gesetzt ist und
SNMP in Enable freigegeben ist."
-- NMS trap annotation
--#TYPE "Thermo-Hygrometer Alarm 5"
--#SUMMARY "Alarm 5: %s"
--#ARGUMENTS {0}
--#SEVERITY INFORMATIONAL
--#TIMEINDEX 99
--#STATE OPERATIONAL
::= 35
wtWebGraphThermHygroAlert6 TRAP-TYPE
ENTERPRISE wtWebGraphThermHygro
VARIABLES { wtWebGraphThermHygroAlarmTrapText }
DESCRIPTION
"Dieser SNMP-Trap wird generiert, wenn der Alarm 6 ausgel<65>st wurde,
das SNMP-Trap-Enable gesetzt ist und
SNMP in Enable freigegeben ist."
-- NMS trap annotation
--#TYPE "Thermo-Hygrometer Alarm 6"
--#SUMMARY "Alarm 6: %s"
--#ARGUMENTS {0}
--#SEVERITY INFORMATIONAL
--#TIMEINDEX 99
--#STATE OPERATIONAL
::= 36
wtWebGraphThermHygroAlert7 TRAP-TYPE
ENTERPRISE wtWebGraphThermHygro
VARIABLES { wtWebGraphThermHygroAlarmTrapText }
DESCRIPTION
"Dieser SNMP-Trap wird generiert, wenn der Alarm 7 ausgel<65>st wurde,
das SNMP-Trap-Enable gesetzt ist und
SNMP in Enable freigegeben ist."
-- NMS trap annotation
--#TYPE "Thermo-Hygrometer Alarm 7"
--#SUMMARY "Alarm 7: %s"
--#ARGUMENTS {0}
--#SEVERITY INFORMATIONAL
--#TIMEINDEX 99
--#STATE OPERATIONAL
::= 37
wtWebGraphThermHygroAlert8 TRAP-TYPE
ENTERPRISE wtWebGraphThermHygro
VARIABLES { wtWebGraphThermHygroAlarmTrapText }
DESCRIPTION
"Dieser SNMP-Trap wird generiert, wenn der Alarm 8 ausgel<65>st wurde,
das SNMP-Trap-Enable gesetzt ist und
SNMP in Enable freigegeben ist."
-- NMS trap annotation
--#TYPE "Thermo-Hygrometer Alarm 8"
--#SUMMARY "Alarm 8: %s"
--#ARGUMENTS {0}
--#SEVERITY INFORMATIONAL
--#TIMEINDEX 99
--#STATE OPERATIONAL
::= 38
wtWebGraphThermHygroAlert9 TRAP-TYPE
ENTERPRISE wtWebGraphThermHygro
VARIABLES { wtWebGraphThermHygroAlarmClearTrapText }
DESCRIPTION
"Dieser SNMP-Trap wird generiert, wenn der Alarm 1 gel<65>scht wurde,
das SNMP-Trap-Enable gesetzt ist und
SNMP in Enable freigegeben ist."
-- NMS trap annotation
--#TYPE "Thermo-Hygrometer Release Alarm 1"
--#SUMMARY "Release Alarm 1: %s"
--#ARGUMENTS {0}
--#SEVERITY INFORMATIONAL
--#TIMEINDEX 99
--#STATE OPERATIONAL
::= 91
wtWebGraphThermHygroAlert10 TRAP-TYPE
ENTERPRISE wtWebGraphThermHygro
VARIABLES { wtWebGraphThermHygroAlarmClearTrapText }
DESCRIPTION
"Dieser SNMP-Trap wird generiert, wenn der Alarm 2 gel<65>scht wurde,
das SNMP-Trap-Enable gesetzt ist und
SNMP in Enable freigegeben ist."
-- NMS trap annotation
--#TYPE "Thermo-Hygrometer Release Alarm 2"
--#SUMMARY "Release Alarm 2: %s"
--#ARGUMENTS {0}
--#SEVERITY INFORMATIONAL
--#TIMEINDEX 99
--#STATE OPERATIONAL
::= 92
wtWebGraphThermHygroAlert11 TRAP-TYPE
ENTERPRISE wtWebGraphThermHygro
VARIABLES { wtWebGraphThermHygroAlarmClearTrapText }
DESCRIPTION
"Dieser SNMP-Trap wird generiert, wenn der Alarm 3 gel<65>scht wurde,
das SNMP-Trap-Enable gesetzt ist und
SNMP in Enable freigegeben ist."
-- NMS trap annotation
--#TYPE "Thermo-Hygrometer Release Alarm 3"
--#SUMMARY "Release Alarm 3: %s"
--#ARGUMENTS {0}
--#SEVERITY INFORMATIONAL
--#TIMEINDEX 99
--#STATE OPERATIONAL
::= 93
wtWebGraphThermHygroAlert12 TRAP-TYPE
ENTERPRISE wtWebGraphThermHygro
VARIABLES { wtWebGraphThermHygroAlarmClearTrapText }
DESCRIPTION
"Dieser SNMP-Trap wird generiert, wenn der Alarm 4 gel<65>scht wurde,
das SNMP-Trap-Enable gesetzt ist und
SNMP in Enable freigegeben ist."
-- NMS trap annotation
--#TYPE "Thermo-Hygrometer Release Alarm 4"
--#SUMMARY "Release Alarm 4: %s"
--#ARGUMENTS {0}
--#SEVERITY INFORMATIONAL
--#TIMEINDEX 99
--#STATE OPERATIONAL
::= 94
wtWebGraphThermHygroAlert13 TRAP-TYPE
ENTERPRISE wtWebGraphThermHygro
VARIABLES { wtWebGraphThermHygroAlarmClearTrapText }
DESCRIPTION
"Dieser SNMP-Trap wird generiert, wenn der Alarm 5 gel<65>scht wurde,
das SNMP-Trap-Enable gesetzt ist und
SNMP in Enable freigegeben ist."
-- NMS trap annotation
--#TYPE "Thermo-Hygrometer Release Alarm 5"
--#SUMMARY "Release Alarm 5: %s"
--#ARGUMENTS {0}
--#SEVERITY INFORMATIONAL
--#TIMEINDEX 99
--#STATE OPERATIONAL
::= 95
wtWebGraphThermHygroAlert14 TRAP-TYPE
ENTERPRISE wtWebGraphThermHygro
VARIABLES { wtWebGraphThermHygroAlarmClearTrapText }
DESCRIPTION
"Dieser SNMP-Trap wird generiert, wenn der Alarm 6 gel<65>scht wurde,
das SNMP-Trap-Enable gesetzt ist und
SNMP in Enable freigegeben ist."
-- NMS trap annotation
--#TYPE "Thermo-Hygrometer Release Alarm 6"
--#SUMMARY "Release Alarm 6: %s"
--#ARGUMENTS {0}
--#SEVERITY INFORMATIONAL
--#TIMEINDEX 99
--#STATE OPERATIONAL
::= 96
wtWebGraphThermHygroAlert15 TRAP-TYPE
ENTERPRISE wtWebGraphThermHygro
VARIABLES { wtWebGraphThermHygroAlarmClearTrapText }
DESCRIPTION
"Dieser SNMP-Trap wird generiert, wenn der Alarm 7 gel<65>scht wurde,
das SNMP-Trap-Enable gesetzt ist und
SNMP in Enable freigegeben ist."
-- NMS trap annotation
--#TYPE "Thermo-Hygrometer Release Alarm 7"
--#SUMMARY "Release Alarm 7: %s"
--#ARGUMENTS {0}
--#SEVERITY INFORMATIONAL
--#TIMEINDEX 99
--#STATE OPERATIONAL
::= 97
wtWebGraphThermHygroAlert16 TRAP-TYPE
ENTERPRISE wtWebGraphThermHygro
VARIABLES { wtWebGraphThermHygroAlarmClearTrapText }
DESCRIPTION
"Dieser SNMP-Trap wird generiert, wenn der Alarm 8 gel<65>scht wurde,
das SNMP-Trap-Enable gesetzt ist und
SNMP in Enable freigegeben ist."
-- NMS trap annotation
--#TYPE "Thermo-Hygrometer Release Alarm 8"
--#SUMMARY "Release Alarm 8: %s"
--#ARGUMENTS {0}
--#SEVERITY INFORMATIONAL
--#TIMEINDEX 99
--#STATE OPERATIONAL
::= 98
wtWebGraphThermHygroAlertDiag TRAP-TYPE
ENTERPRISE wtWebGraphThermHygro
VARIABLES {
wtWebGraphThermHygroDiagErrorIndex,
wtWebGraphThermHygroDiagErrorMessage
}
DESCRIPTION
"Dieser SNMP-Trap wird generiert, wenn eine Diagnose/Error-Message ausgel<65>st wurde,
das SNMP-Trap-Enable gesetzt ist und
SNMP in Enable freigegeben ist."
-- NMS trap annotation
--#TYPE "Thermo-Hygrometer Diagnose Error"
--#SUMMARY "Diagnose error index: %d, Message: %s."
--#ARGUMENTS {0,1}
--#SEVERITY MAJOR
--#TIMEINDEX 99
--#STATE DEGRADED
::= 110
END